首页> 外文期刊>Journal of orthopaedic research >Evaluation of Heat Generation in Unidirectional Versus Oscillatory Modes During K‐Wire Insertion in Bone
【24h】

Evaluation of Heat Generation in Unidirectional Versus Oscillatory Modes During K‐Wire Insertion in Bone

机译:在骨中k线插入期间单向与振荡模式中的热产生的热发电

获取原文
获取原文并翻译 | 示例
           

摘要

ABSTRACT Heat generation during insertion of Kirschner wires (Kwires) may lead to thermal osteonecrosis and can affect the construct fixation. Unidirectional and oscillatory drilling modes are options for Kwire insertion, but understanding of the difference in heat generation between the two modes is lacking. The goal of this study was to compare the temperature rise during Kwire insertion under these two modes and provide technical guidelines for Kwire placement to minimize thermal injury. Ten orthopedic surgeons were instructed to drill holes on hydrated ex vivo bovine bones under two modes. The drilling trials were evaluated in terms of temperature, thrust force, torque, drilling time, and tool wear. The analysis of variance showed that the oscillatory mode generated significantly lowered peak bone temperature rise (13% lower mean value,p = 0.036) over significantly longer drilling time (46% higher mean time,p 0.001) than the unidirectional mode. Drilling time had significant effect on peak bone temperature rise under both modes (p 0.001) and impact of peak thrust force was significant under oscillatory mode (p 0.001). These findings suggest that the drilling mode choice is a compromise between peak temperature and bone exposure time. Shortening the drilling time was the key under both modes to minimize temperature rise and thermal necrosis risk. To achieve faster drilling, technique analysis found that shaky and intermittent drilling with moderate thrust force are preferred techniques by small vibration of the drill about the Kwire axis and slight liftup of the Kwire once or twice during drilling. ?2019 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 37:19031909, 2019
机译:摘要插入Kirschner电线(Kwires)期间的发热可能导致热骨折,可以影响构造固定。单向和振荡钻探模式是Kwire插入的选择,但是缺乏对两种模式之间的发热差异的理解。本研究的目的是将Kwire插入期间的温度升高进行比较,并为kwire放置提供技术指南,以最大限度地减少热损伤。指示十个骨科外科医生在两种模式下指示钻孔钻孔钻孔。在温度,推力,扭矩,钻井时间和工具磨损方面评估钻井试验。方差分析表明,振荡模式产生显着降低的峰值骨温度(较低平均值,P = 0.036)显着更长的钻井时间(平均时间为46%,P <0.001)而不是单向模式。钻井时间对两种模式(P <0.001)下峰值骨温度升高产生显着影响,并且在振荡模式下峰值推力的影响是显着的(P <0.001)。这些发现表明,钻井模式选择是峰值温度和骨暴露时间之间的折衷。缩短钻井时间是两种模式下的关键,以最大限度地减少温度上升和热坏死风险。为了实现更快的钻井,技术分析发现,具有中等推力的摇晃和间歇钻孔是优选的技术,通过钻孔轴的小振动以及钻孔一次或两次kwire的轻微升力。 ?2019年骨科研究会。由Wiley Hearyicals,Inc.J orthop Res 37:19031909,2019出版

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号