首页> 外文期刊>Journal of mechanics in medicine and biology >BIOMECHANICAL STUDY ON MECHANISMS OF FEMORAL NECK FRACTURE AFTER HIP RESURFACING ARTHROPLASTY
【24h】

BIOMECHANICAL STUDY ON MECHANISMS OF FEMORAL NECK FRACTURE AFTER HIP RESURFACING ARTHROPLASTY

机译:髋关节置换术后股骨颈骨折机制的生物力学研究

获取原文
获取原文并翻译 | 示例
           

摘要

Hip resurfacing arthroplasty (HRA) is a long-established procedure. It is a minimally invasive surgery where the surgical wound is relatively small to facilitate a shorter recovery period. HRA remained a popular option among the patients allowing better range of motion of the joint compared to that of total hip arthroplasty (THA). Although HRA is associated with the above advantages, complications involving femoral neck fractures after surgery still occur. Therefore, the present study attempts to assess the impact of stress under various alignment conditions and different scenarios in surgical errors upon the femoral neck in hip resurfacing prostheses (HRP) that may be encountered during the procedure using finite element analysis (FEA) technique. The results showed that anteversion implantation errors on femoral components should be avoided, and that the main reason that causes femoral neck fracture is related to the stress shielding effect generated internally in the femoral neck. Methods to prevent the incidence of such events are a major obstacle to be solved in the future.
机译:HIP Resurfacing关节造身术(HRA)是一项良好的程序。它是一种微创手术,其外科伤口相对较小,便于较短的恢复期。与总髋关节置换术(THA)相比,HRA在患者中仍然是患者的流行选择,允许联合的更好运动范围。尽管HRA与上述优点有关,但仍发生手术后患有股骨颈骨折的并发症。因此,本研究试图评估各种对准条件下应力的影响以及在使用有限元分析(FEA)技术的过程中可能遇到的髋关节重叠假体(HRP)中的股骨颈上的手术误差的不同情景。结果表明,应避免对股骨部件上的抗病植入误差,并导致股骨颈骨折的主要原因与股骨颈内部产生的应力屏蔽效果有关。防止此类事件发生的方法是未来解决的主要障碍。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号