首页> 外文期刊>Journal of Medical Devices >Continuous, Real-Time, Noninvasive Hemodynamic Cardiac Doppler Monitoring With a Novel Hands-Free Device: A Feasibility Study Compared to Standard Echo
【24h】

Continuous, Real-Time, Noninvasive Hemodynamic Cardiac Doppler Monitoring With a Novel Hands-Free Device: A Feasibility Study Compared to Standard Echo

机译:连续,实时,非血液动力学心脏多普勒监测用新型免提装置:与标准回声相比的可行性研究

获取原文
获取原文并翻译 | 示例

摘要

This study sought to determine the feasibility of using noninvasive cardiac hemodynamics (NICHE), a new noninvasive Doppler-based device, to monitor real-time, simultaneous tissue and blood-flow Doppler measurements in a clinical setting, and to obtain preliminary performance data compared to a commercially available system. Doppler-based measurements have been shown to correlate well with invasive hemodynamic data and diastolic function, but their use in clinical applications has been limited by various technical issues. The NICHE device was developed to obtain simultaneous tissue and blood-flow Doppler measurements automatically, in real-time and in a hands-free manner. Thirty participants (ten normal volunteers and 20 patients in a cardiac rehab program) underwent standard echocardiographic/Doppler studies followed immediately by NICHE monitoring. Early diastolic transmitral blood-flow velocity (E) and tissue Doppler myocardial wall velocity during early relaxation (E') were acquired using a standard echo device; and E/E' was derived post hoc. NICHE measurements included E, E', and directly measured instantaneous E/E'. NICHE was successfully used in 28 participants. Measurements of E-NICHE ranged from 40 cm/s to over 120 cm/s and correlated well with E-echo (R = 0.93). E'(NICHE) ranged from 2 to 23 cm/s and correlated well with the averaged E'(echo) (R = 0.91). Directly measured E/E'(NICHE) ratios ranged from 3 to 23 and correlated well with derived E/E'(echo) (R = 0.91). The NICHE device can monitor patients in a hands-free manner and can supply real-time Doppler derived measurements of hemodynamic parameters and diastolic function that correlate well with measurements from standard echo devices.
机译:本研究试图确定使用非侵入性心脏血流动力学(NICHE),一种新的非侵入性多普勒基装置的可行性,以监测临床环境中的实时,同时组织和血流多普勒测量,并获得初步性能数据到市售的系统。已经显示了基于多普勒的测量与侵入性血液动力学数据和舒张功能良好,但它们在临床应用中的使用受到各种技术问题的限制。开发了利基装置,以获得同时组织和血流多普勒测量,实时和免提。三十名参与者(十个正常志愿者和20名心脏康复程序患者)接受了标准超声心动图/多普勒研究,然后立即通过利基监测。利用标准回声装置获得早期弛豫期间的早期舒张式传递血流速度(E)和组织多普勒心肌壁速度;和e / e'派生后HOC。利基测量包括E,E',直接测量瞬时E / e'。利基成功用于28名参与者。 E-NICHE的测量范围为40cm / s至超过120cm / s,并与电子回波相关(r = 0.93)。 e'(niche)的范围为2至23cm / s,并且平均e'(回声)(r = 0.91)相关。直接测量的E / E'(NICHE)比率范围为3至23并与衍生的E / E'(回声)(R = 0.91)相关。利基装置可以免提监测患者,可以免提,可以提供血液动力学参数和舒张功能的实时多普勒衍生测量,舒张函数与标准回波器件的测量相比好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号