[~(18)F]ML10 is a promising novel low molecular weight positron emission tomography probe for apoptosis. As part of the quality control to support clinical studies for cancer therapy monitoring in the GSK Clinical Imaging Centre, a simple and sensitive liquid chromatography mass spectrometry method has been developed and validated for the quantification of total ML10 and impurity content in the final product. Chromatographic separation of ML10 and its radiolabelling precursor and impurities was achieved. Mass curves were constructed from a concentration range of ML10 and known impurities and were linear. Quantification was achieved by comparison of the area under the curve for ML10 content (m/z = 205) and the mass curve. The method was validated over a concentration range of 0.1-1mug/ml.
机译:[〜(18)F] ML10是一种有前途的新型低分子量正电子发射断层扫描探针,用于凋亡。 作为支持GSK临床成像中心癌症治疗监测临床研究的质量控制的一部分,已经开发了简单敏感的液相色谱质谱法,并验证了最终产品中总ML10和杂质含量的定量。 实现了ML10的色谱分离及其放射性标记前体和杂质。 质量曲线由ML10和已知杂质的浓度范围构成,并且是线性的。 通过比较ML10含量(M / Z = 205)和质量曲线曲线下的面积来实现量化。 该方法在0.1-1Mug / mL的浓度范围内验证。