A point of application study to determine the accuracy, precision and reliability of a low-cost balance plate for center of pressure measurement

机译:一种应用研究点,以确定压力测量中心低成本平衡板的准确性,精度和可靠性

获取原文
获取原文并翻译 | 示例
           

摘要

Changes in postural sway measured via force plate center of pressure have been associated with many aspects of human motor ability. A previous study validated the accuracy and precision of a relatively new, low-cost and portable force plate called the Balance Tracking System (BTrackS). This work compared a laboratory-grade force plate versus BTrackS during human-like dynamic sway conditions generated by an inverted pendulum device. The present study sought to extend previous validation attempts for BTrackS using a more traditional point of application (POA) approach. Computer numerical control (CNC) guided application of similar to 155 N of force was applied five times to each of 21 points on five different BTrackS Balance Plate (BBP) devices with a hex-nose plunger. Results showed excellent agreement (ICC 0.999) between the POAs and measured COP by the BBP devices, as well as high accuracy (1% average percent error) and precision (0.1 cm average standard deviation of residuals). The ICC between BBP devices was exceptionally high (ICC 0.999) providing evidence of almost perfect inter-device reliability. Taken together, these results provide an important, static corollary to the previously obtained dynamic COP results from inverted pendulum testing of the BBP. (C) 2018 Elsevier Ltd. All rights reserved.
机译:通过力板压力测量的姿势摇摆的变化已经与人类电机能力的许多方面有关。先前的研究验证了一种称为平衡跟踪系统(Btracks)的相对较新的低成本和便携式力板的准确性和精度。这项工作比较了由倒置摆在倒置摆片产生的人类动态摇摆条件下的实验室级力板与Btracks。目前的研究试图使用更传统的应用点(POA)方法来扩展以前的BTRAcks验证尝试。计算机数值控制(CNC)类似于155 n力的引导施用是用六角柱塞的五个不同BTRACKS平板(BBP)器件上的21个点中的5倍。结果表明,POA之间的优异协议(ICC> 0.999)通过BBP设备测量COP,以及高精度(误差)和精度(& 0.1cm的差异标准偏差)。 BBP设备之间的ICC非常高(ICC> 0.999)提供几乎完美的内部可靠性的证据。总之,这些结果为先前获得的动态COP提供了一个重要的静态必然原因来自BBP的倒立摆检测。 (c)2018年elestvier有限公司保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号