The development and performance characterization of a new differential pressure-based flow meter for human breath measurements is presented in this article. The device, called a Confined Pitot Tube, is comprised of a pipe with an elliptically shaped expansion cavity located in the pipe center, and an elliptical disk inside the expansion cavity. The elliptical disk, named Pitot Tube, is exchangeable, and has different diameters, which are smaller than the diameter of the elliptical cavity. The gap between the disk and the cavity allows the flow of human breath to pass through. The disk causes an obstruction in the flow inside the pipe, but the elliptical cavity provides an expansion for the flow to circulate around the disk, decreasing the overall flow resistance. We characterize the new sensor flow experimentally and theoretically, using Comsol Multiphysics((R)) software with laminar and turbulent models. We also validate the sensor, using inhalation and exhalation tests and a reference method. (c) 2016 American Institute of Chemical Engineers AIChE J, 62: 956-964, 2016
机译:本文介绍了一种新的基于压差的人体呼吸测量流量计的开发和性能表征。该设备被称为“受限皮托管”,由一个在管道中心处具有椭圆形膨胀腔的管道和一个在膨胀腔内部的椭圆形圆盘组成。椭圆盘(称为皮托管)可以互换,并且具有不同的直径,这些直径小于椭圆腔的直径。盘和腔之间的间隙允许人的呼吸通过。圆盘会阻碍管道内部的流动,但椭圆形腔体会为流在圆盘周围循环提供膨胀,从而降低了总体流动阻力。我们使用带有层流模型和湍流模型的Comsol Multiphysics(R)软件,在实验和理论上表征了新传感器的流量。我们还使用吸气和呼气测试以及参考方法来验证传感器。 (c)2016美国化学工程师学会AIChE J,62:956-964,2016

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号