Does an accelerometer accurately measure steps taken under controlled conditions in adults with mild multiple sclerosis?

机译:加速度计是否准确地测量在具有轻度多发性硬化的成人的受控条件下采取的步骤?

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Accurate measurement is required by researchers and clinicians who are interested in the physical activity behavior of individuals with multiple sclerosis (MS). Advances in technology have resulted in an increased number of motion sensors such as pedometers and accelerometers that are worn on the body and that measure bodily movement. Accelerometers are becoming less expensive and more user-friendly, but there is limited evidence regarding the accuracy of measurement in persons with MS. OBJECTIVE: The present study examined the accuracy of an ActiGraph accelerometer for measuring steps taken during controlled conditions in persons with MS compared with a sample of individuals without MS. METHODS: The participants were 24 adults with mild MS and 24 adults without MS who undertook three 6-minute periods of walking at 54, 80, and 107 m.min(-1) on a motor-driven treadmill. We measured steps taken through observation and an ActiGraph model 7164 accelerometer worn around the waist above the right hip. RESULTS: The accelerometer accurately measured steps during moderate (80 m.min(-1)) and fast (107 m.min(-1)) walking in both persons with MS and control subjects. There was a small degree of underestimation of step counts ( approximately 4% error) for the accelerometer during slower walking (54 m.min(-1)) in both persons with MS and control subjects. CONCLUSIONS: Such findings support the accuracy of a waist worn ActiGraph accelerometer for the measurement of steps in persons with MS and control subjects.
机译:背景:对具有多种硬化症(MS)的个体的身体活动行为感兴趣的研究人员和临床医生需要准确测量。技术的进步导致了诸如在身体上佩戴的测量仪和加速度计的运动传感器增加,并且测量身体运动。加速度计变得不那么昂贵,更用户友好,但有限的证据证明了与MS的人员的测量准确性。目的:本研究检测了用于测量在与MS的受控条件下拍摄的步骤测量步骤的ACTIGRAPH加速度计的准确性与没有MS的个体样本。方法:参与者是24名成人,其中24名同性恋MIDS和24位成年人,如果在电动跑步机上进行三个6分钟的行走,6分钟,80.7毫米,则在54,80和107 m.in(-1)上。我们通过观察和围绕右臀部的腰部磨损的侧面磨损的散射型7164加速度计进行了测量的步骤。结果:加速度计在中等(80 m.Min(-1))和快速(107 m.in(-1))中携带MS和控制受试者的人的快速测量步骤。在MS和控制主题的人员中,在较慢的步行(54m.min(-1))较慢的步行(54m.min(-1))中,对加速度计的步骤计数(约4%误差)有很小程度的低估。结论:这种研究结果支持腰部磨损的采用仪的准确性,用于测量与MS和控制受试者的人员中的步骤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号