首页> 外文期刊>Magma: Magnetic resonance materials in physics, biology, and medicine >Reduction of Ghost Artifacts in ~(31)P EPSI Using a Gradient Impulse Response Model
【24h】

Reduction of Ghost Artifacts in ~(31)P EPSI Using a Gradient Impulse Response Model

机译:使用梯度脉冲响应模型减少〜(31)P EPSI中的幽灵伪影

获取原文
获取原文并翻译 | 示例
           

摘要

Accelerated magnetic resonance spectroscopic imaging (MRSI) sequences like echo-planar spectroscopic imaging (EPSI) place high demands on the MR gradient system. Imperfections in gradient trajectories lead to ghost artifacts in conventionally reconstructed MRSI data due to phase inconsistencies between individual gradient echoes. The separate phase-corrected reconstruction of odd and even echo trains avoids these ghosts; however, this approach limits the spectral bandwidth, which is a disadvantage for ~(31)P MRSI at B_0 = 7T.
机译:加速磁共振光谱成像(MRSI)序列,如回波平面光谱成像(EPSI)对MR梯度系统提供高要求。 由于各个梯度回波之间的相位不一致,梯度轨迹中的缺陷导致幽灵伪像在传统上重建的MRSI数据中。 单独的相位校正的奇数和甚至回声列车的重建避免了这些幽灵; 然而,该方法限制了光谱带宽,这是B_0 = 7T的〜(31)P MRSI的缺点。