In-vitro wear simulation of sequentially crosslinked and annealed polyethylene acetabular liners: 14 years of results

机译:依次交联和退火的聚乙烯髋臼衬里的体外磨损模拟:14年的结果

获取原文
获取原文并翻译 | 示例
           

摘要

Sequentially crosslinked and annealed polyethylene (SXL) is a second generation highly crosslinked polyethylene that has been available for joint replacement with clinical success. The purpose of this study is to characterize SXL cups through an analysis of almost 14 years of historical in vitro wear simulation data. Three characteristics were evaluated in this study (1) Wear rate, (2) linear penetration and (3) shelf life effect on wear.
机译:依次交联和退火的聚乙烯(SXL)是第二代高度交联的聚乙烯,其可用于临床成功的关节置换。 本研究的目的是通过分析近14年的历史体外磨损模拟数据来表征SXL杯。 本研究评估了三种特征(1)磨损率,(2)线性穿透和(3)磨损的保质期效果。