Comparison of invasive and non-invasive blood pressure monitoring during clinical anaesthesia in dogs

机译:犬临床麻醉期间侵袭性和非侵袭性血压监测的比较

获取原文
获取原文并翻译 | 示例
       

摘要

Monitoring blood pressure during anaesthesia is widely recommended in man and animals. The accuracy of any device used to measure blood pressure is an important consideration when selecting monitoring equipment, the ANSI/AAMI SP10 standard is widely cited in this respect in recent veterinary publications. Blood pressure was monitored using invasive and non-invasive techniques during clinical anaesthesia in 19 dogs. The results were compared using Bland-Altman analysis. The bias (and limits of agreement) between invasive and non-invasive measurement was 7.1 mmHg (+/34.7) for systolic blood pressure, 1.8 mmHg (+/27.4) for mean blood pressure and 6.9 mmHg (+/27.5) for diastolic blood pressure. In a clinical setting the bias between invasive and non-invasive measurement techniques was similar or smaller than laboratory reports, however the limits of agreement were considerably wider suggesting that care should be exercised when interpreting NIBP values.
机译:在人和动物中,广泛推荐在麻醉期间监测血压。 用于测量血压的任何装置的准确性是选择监测设备时的重要考虑因素,在近期兽医出版物中广泛引用ANSI / AAMI SP10标准。 使用19只犬的临床麻醉期间使用侵入性和非侵入性技术监测血压。 使用Bland-Altman分析进行比较结果。 侵袭性和非侵入性测量之间的偏差(和协议限制)为17.1mmHg(+ / 34.7),用于收缩压,1.8mmHg(+ / 27.4),用于平均血压和6.9mmHg(+ / 27.5)用于舒张血 压力。 在临床环境中,侵入性和非侵入性测量技术之间的偏差相似或小于实验室报告,但是一致性的限制较宽,表明在解释NIBP值时应该注意。