The reliability and variation of acetabular component anteversion measurements from cross-table lateral radiographs.

机译:跨桌横射射线照相的髋臼分量安踏测量的可靠性和变化。

获取原文
获取原文并翻译 | 示例
       

摘要

Although cross-table lateral (CL) radiographs are frequently used to assess acetabular component anteversion, the reliability of this method has not been established. We compared serial CL radiographs with computed tomography (CT) scans for 98 total hip arthroplasty patients (119 hips) undergoing surveillance of primary or revision total hip arthroplasty. Acetabular anteversion averaged 26.1 degrees (range, -2 degrees to 48.3 degrees ) on CL imaging and 28.8 degrees (range, -7 degrees to 54 degrees ) on CT scan. There was a strong correlation between anteversion determined from CT scans and serial CL images. However, variation on serial CL studies exceeded 10 degrees for 20% of patients. Although CL imaging provides acceptable assessment of general component position, it has limited use for precise analysis in research, outcome reporting, or determination of cause of implant failure.
机译:虽然跨表横向(CL)X光片经常用于评估髋臼部件的安踏,但尚未建立该方法的可靠性。 我们将串行CL XcistProve与计算机断层扫描(CT)扫描进行比较98个总髋关节置换术患者(119髋)进行初级或修订总髋关节置换术的监测。 CT扫描上的髋臼安置平均为CL成像和28.8度(范围,-7度至54度)的26.1度(范围,-2度至48.3度)。 从CT扫描和串行CL图像确定的反率之间存在强烈的相关性。 然而,患者的序列CL研究的变化超过了10%的患者。 虽然CL成像提供了一般组分位置可接受的评估,但它有限用于研究,结果报告或植入原因的研究原因。