Evaluation of the Sterility of Single-Use Vials Undergoing Multiple Access Following Application of a Closed System Transfer Device

机译:在闭合系统转移装置施加后进行多次访问的单用小瓶的无菌评价

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Background: Closed system transfer devices (CSTD) are designed for safe handling of hazardous drugs from preparation to administration. According to NIOSH, these devices are airtight and leak-proof. While these devices protect staff, as a closed system they could also minimize microbial contamination. Objective: To test whether attaching a CSTD (Equashield?) to single-use vials can minimize microbial contamination and extend the "use-by" date following multiple withdrawals under extreme-use-conditions. Methods: An Equashield? vial adapter was attached to three 20 mL vials (A, B, C) containing sterile TSB growth medium and placed in each of 6 biological safety cabinets weekly for 16 weeks. Vial A (control) had no medium removed during the week. One mL of medium was removed once daily x5 days from vial B, and twice daily x5 days from vial C. At day 5, vials were collected, incubated at 37°C for 14 days and inspected visually every 2 days for microbial growth. As a positive control, TSB vials were inoculated with less than 102 of 5. epidermidis ATC 12228. As a negative control, an unopened vial of TSB was incubated for the duration of the study. Results: All positive control vials demonstrated growth within 48 hours. All negative control vials showed no growth throughout the study. During the 16-week study all accessed vials remained sterile following storage at room temperature for 5 days and subsequent incubation for 14 days. None of the 192 vials accessed 1440 times or the 96 vials that had the CSTD attached but had no broth removed demonstrated contamination. The 95% confidence interval of the contamination rate is 0.000 to 0.035%. Conclusions: Attachment of a CSTD adapter to single-use vials within an ISO-5 environment has the ability to maintain sterility following multiple withdrawals during 5 days and stored in worse than ISO-5 conditions.
机译:背景:闭合系统转移装置(CSTD)设计用于安全地处理危险药物从制备到给药。根据Niosh的说法,这些装置是密闭和防漏。虽然这些设备保护员工,但作为一个封闭的系统,它们也可以最大限度地减少微生物污染。目的:要测试是否将CSTD(QuaShield?)连接到单使用小瓶可以最小化微生物污染,并在极端使用条件下进行多次取款后延长“使用”日期。方法:赤架吗?将样瓶适配器连接到含有无菌TSB生长培养基的三种20ml小瓶(A,B,C),并每周置于6个生物安全柜中的每一个中16周。小瓶A(对照)本周没有染色培养基。每天从小瓶B中每天X5天移除一毫升培养基,每天X5天从小瓶C.在第5天,收集小瓶,在37℃温育14天,每2天在37℃温育14天,每2天检查一次微生物生长。作为阳性对照,用小于102的5. ePIDermidis ATC 12228接种TSB小瓶。作为阴性对照,在研究期间孵育未开封的TSB的花瓶。结果:所有阳性对照小瓶均在48小时内显示出增长。所有阴性对照小瓶都在整个研究中没有显示出增长。在16周的研究中,所有进入的小瓶在室温下储存后仍然无菌5天,随后孵育14天。 192瓶的192瓶中没有一次,其中包含CSTD的96个小瓶,但没有肉汤被移除污染。污染率的95%置信区间为0.000至0.035%。结论:将CSTD适配器与ISO-5环境中的单使用小瓶的附着能够在5天内进行多次取液后保持无菌,并在比ISO-5条件储存更差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号