A new injection rate estimation technique for on-site screening test of medication infusion pump by nurses

机译:护士用药输液泵现场筛选试验的新注射率估算技术

获取原文
获取原文并翻译 | 示例
           

摘要

Medication infusion pumps are the most popular device in almost all areas of a hospital; therefore, it is important to frequently inspect the accuracy of the infusion pump operation to prevent underdose/overdose accidents. However, the conventional infusion pump inspection devices are not suitable for quick and convenient on-site inspection by nurses. In this study, a new IR estimation technique for peristaltic infusion pumps that facilitates on-site pre-screening test with shorter inspection time was proposed. A thin membrane potentiometer was attached to a catheter and the actual IR was estimated based on a time interval between two successive line pushes of an identical cam follower using power function estimation. To evaluate the performance of the proposed IR estimation technique, in vitro experiments were performed using 11 infusion pumps (three for Infusion Pump SET 1 (IPSET-1) and eight for Infusion Pump SET 2 (IPSET-2)) with the same model. In experiments, error rate between the actual and the measured values (using conventional inspection device) were 0.04-1.17% range for IPSET-1 and 2.09-4.32% for IPSET-2, and those between the actual and the estimated values (using proposed method) were 0.02-0.62% range for IPSET-1 and 1.31-4.23% for IPSET-2. The proposed technique had almost equivalent performance with a commercial inspection device, but the time for inspection was reduced to almost one third. We expect that the proposed technique can provide a tool for simple and convenient on-site pre-screening of infusion pumps by nurses to improve patient safety.
机译:药物输液泵是几乎所有医院领域最受欢迎的设备;因此,重要的是要经常检查输液泵操作的准确性,以防止耗能/过量的事故。然而,传统的输液泵检查装置不适合护士快速方便的现场检查。在该研究中,提出了一种新的IR估计技术,用于促进具有较短检查时间的促进现场预筛选试验的蠕动输注泵。将薄膜电位器连接到导管上,并且基于使用功率函数估计的两个连续线推动相同的凸轮从动件之间的时间间隔估计实际的IR。为了评估所提出的IR估计技术的性能,使用具有相同型号的11个输液泵(用于输液泵组1(IPSet -1)和八个用于输液泵组2(IPset -2))进行体外实验。在实验中,IPSet-2的IPSet-1和2.09-4.32%的实际值和测量值之间的错误率为0.04-1.17%,以及实际和估计值之间的值(使用所提出的)方法)IPSet-1的IPSet-1和1.31-4.23%的0.02-0.62%。该技术几乎具有商业检测装置的等效性能,但检查的时间降至近三分之一。我们预计所提出的技术可以为护士提供简单方便的现场预筛分的工具,以提高患者安全性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号