A Novel Approach to Assess Performance of Glucometer: Error Grid Analysis

机译:一种评估血糖仪性能的新方法:误差网分析

获取原文
获取原文并翻译 | 示例
           

摘要

Glucose meters are excellent tools for self-monitoring of blood glucose especially in the circumstances where continuous monitoring is mandatory and at decision-making levels. Tight glycemic control protocols are important for preventing ill effects of fluctuating glucose levels. As technology advances, glucometers are getting better in terms of quality of results. The purpose here is to study the quality of glucose meter results in terms of clinical outcomes. The study was conducted in a tertiary care referral hospital. A total of 125 patients were recruited from pediatric wards. Bland-Altman plot, Parke error grid, and surveillance error grid analysis were used for comparing the results of the glucose meters with a standard laboratory method. It was found that there was a significant difference between the results of the 2 methods. Although usually minimal, the glucose meter results deviated from the results of the standard laboratory method. This will affect overall patient care especially in emergency conditions. Sometimes, the risk may be so high that the patient may be mislabeled as hypoglycemic when they actually are hyperglycemic and vice versa. This study is the first of its kind because no similar studies have been reported in the pediatric population. For effective use of glucose meters, the devices should give as accurate results as possible. Results should not only be accurate but also precise, without which critical errors may be possible. We recommend that for any glucose meter there should be regular calibration so that agreement with the reference laboratory method is maintained and effective medical decisions are possible.
机译:葡萄糖仪是用于血糖的自我监测的优秀工具,特别是在连续监测是强制性和决策水平的情况下。紧张的血糖控制方案对于防止葡萄糖水平的不良影响很重要。随着技术的进步,刺激仪在结果质量方面越来越好。这里的目的是研究葡萄糖仪的质量导致临床结果。该研究是在高等护理推荐医院进行的。共有125名患者从儿科病房招募。 Bland-Altman Plot,PARKE错误网格和监控误差电网分析用于比较葡萄糖仪表的标准实验室方法。发现2种方法的结果之间存在显着差异。虽然通常是最小的,但葡萄糖仪表结果偏离了标准实验室方法的结果。这将影响整体患者护理,特别是在紧急情况下。有时,风险可能如此之高,因为当他们实际是高血糖时,患者可能被误导为降血糖,反之亦然。本研究是其第一个,因为在儿科人口中没有报道类似的研究。有效使用葡萄糖仪,这些器件应尽可能准确地给出。结果不仅应该准确,而且还可以精确,无论哪些严重错误都可能是可能的。我们建议,对于任何葡萄糖仪,应该定期校准,以维持与参考实验室方法的协议,有效的医学决策是可能的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号