The characteristics of scintillation counters measured in a 90-MeV proton beam have been presented. Two-meter long scintillators BC-404 and BC-408, which were viewed from both ends with photoelectron multipliers R1828-01 (PMT) have been used. Both the amplitude and time distributions of the PMT signals are well described by the Gaussian function. It turned out that the amplitude resolution sigma(A)/A weakly depends on the type of scintillator and the proton position along the counter, and lies in the interval between 4 and 5%. The best time resolution sigma(T) is about 50 ps near the counter ends and 63 ps at the center.
机译:已经介绍了在90meV质子束中测量的闪烁计数器的特性。 已经使用了从两端观察的两米长的闪烁体BC-404和BC-408已经使用了光电子倍增器R1828-01(PMT)。 PMT信号的幅度和时间分布在高斯函数很好地描述。 事实证明,幅度分辨率Sigma(a)/ a弱取决于沿着计数器的闪烁体和质子位置的类型,并且在4到5%之间的间隔。 最佳时间分辨率Sigma(t)在计数器端附近约50 p,中心在中心处为63 ps。

著录项

  • 来源
    《Physics of particles and nuclei》 |2018年第1期|共5页
  • 作者单位

    Kurchatov Inst Natl Res Ctr Inst High Energy Phys Protvino 142281 Moscow Oblast Russia;

    Kurchatov Inst Natl Res Ctr Inst High Energy Phys Protvino 142281 Moscow Oblast Russia;

    Kurchatov Inst Natl Res Ctr Inst High Energy Phys Protvino 142281 Moscow Oblast Russia;

    Kurchatov Inst Natl Res Ctr Inst High Energy Phys Protvino 142281 Moscow Oblast Russia;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 粒子物理学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号