In this study a new electrochemical histamine sensor was proposed by using Ni-based metalorganic framework (Ni-BTC, BTC = 1,3,5-benzenetricarboxylate) crystals and multi-walled carbon nanotubes modified glassy carbon electrode. The modified electrode exhibited excellent electrocatalytic activity for the electro-oxidation of histamine with relatively high sensitivity and stability. The electro-oxidation of histamine was irreversible and exhibited an absorption-controlled behavior. A calibration curve for histamine in the concentration range from 1.00 to 160.00 mu M with a detection limit of 0.41 mu M (at signal to noise 3) and sensitivity of 0.19 mu A mu M-1 was obtained. The suggested sensor was successfully used for quantitative determination of histamine in spiked human urine samples with satisfactory results.
机译:在该研究中,通过使用基于Ni的金属有机框架(Ni-BTC,BTC = 1,3,5-苯所述羧酸甲酯)晶体和多壁碳纳米管改性玻璃电极来提出新的电化学组胺传感器。 改性电极表现出具有相对高灵敏度和稳定性的组胺的电氧化的优异的电催化活性。 组胺的电氧化是不可逆的并且表现出吸收控制的行为。 浓度范围为0.41μm(以信号至噪声3)的检测限为1.00至160.00μm的浓度范围的校准曲线和0.19μm-1的灵敏度。 建议的传感器成功地用于定量测定尖刺人尿液中的组胺,结果令人满意。