首页> 外文期刊>Respiration: International Review of Thoracic Diseases >Which Nebulizer Position Should Be Avoided? An Extended Study of Aerosol Delivery and Ventilator Performance during Noninvasive Positive Pressure Ventilation
【24h】

Which Nebulizer Position Should Be Avoided? An Extended Study of Aerosol Delivery and Ventilator Performance during Noninvasive Positive Pressure Ventilation

机译:应该避免哪种雾化器位置? 在非侵入性阳性压力通风中的气溶胶输送和呼吸机性能的扩展研究

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Research on the effect of nebulizer location on aerosol delivery during noninvasive ventilation has reached inconsistent conclusions. Objective: To investigate the effects of nebulizer position on aerosol delivery efficiency and ventilator performance during noninvasive ventilation. Methods: The Active Servo Lung 5000 respiratory simulation system (ASL5000) was used to simulate a COPD patient. The noninvasive ventilator was set to the spontaneous breathing mode. Six nebulizer positions, 2 exhalation valve types (single-arch exhalation port and whisper swivel), 4 combinations of inspiratory and expiratory pressure, and 2 respiratory rates were used. Results: Significant differences between nebulizer positions existed in aerosol delivery ( p p Conclusions: The closer the nebulizer was to the exhalation valves or ventilator, the lower the aerosol delivery efficiency. Nebulizer position had little clinically significant effect on ventilator performance.
机译:背景:在非侵入性通风期间雾化器位置对气溶胶递送的影响的研究已达到不一致的结论。 目的:探讨雾化器位置对非侵入性通气期间气溶胶输送效率和呼吸机性能的影响。 方法:活性伺服肺5000呼吸仿真系统(ASL5000)用于模拟COPD患者。 非侵入式呼吸机设定为自发呼吸模式。 六种雾化器位置,2种呼气阀门类型(单拱呼气口和耳语旋转),4个吸气和呼气压力的组合,使用2个呼吸率。 结果:气溶胶递送中存在的雾化器位置有显着差异(P P P结果:雾化器越接近呼气阀或呼吸机,气溶胶输送效率越低。雾化器位置对呼吸机性能几乎没有临床显着影响。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号