Dosimetric evaluation of the INTRABEAM system for breast intraoperative radiotherapy: A single-institution experience

机译:乳腺术中放射治疗肠内系统的剂量评估:单机制经验

获取原文
获取原文并翻译 | 示例
       

摘要

Breast intraoperative radiotherapy (IORT) with the INTRABEAM system uses a 50 kV x-ray source to deliver a single fraction of radiation therapy to the lumpectomy cavity during breast-conserving surgery. We seek to perform a dosimetric analysis of the lumpectomy cavity for rigid spherical applicators. Water phantom measurements were acquired to validate the vendor-provided x-ray calibration. The planning target volume (PTV) was defined as a 10 mm expansion beyond the spherical applicator, a dose-volume histogram (DVH) was generated and dose-volume parameters [D-min, D-1mm, V-90, V-80, V-50, HI] were reported. Additionally, the therapeutic treatment depth using the 90 and 80% isodose level was computed [R-90, R-80]. When the percent depth dose (PDD) is normalized to the surface of the applicator, smaller applicators have a steeper PDD. For a prescription dose of 20 Gy to the surface of the applicator, the range of dose-volume parameters for the PTV was: 3.15 to 6.84 Gy for D-min, 16.2 to 17.6 Gy for D-1mm, 2.6 to 6.9% for V-90, 5.5 to 15.1% for V-80, and 21.1 to 55.6% for V-50. For applicators 15 to 50 mm in diameter, the reported values were: 6.35 to 2.9 for HI, 0.53 to 0.85 mm for R-90, and 1.18 to 1.85 mm for R-80. Smaller applicators have reduced PTV coverage but elevated HI because the attenuation of the beam proximal to the source is more pronounced. Additionally, the presence of the aluminum filter for small applicators (<= 30 mm) increases PTV coverage but reduces the dose rate on the applicator surface. The delivery of IORT is performed in the OR without the use of image-based planning. To overcome this limitation, we have generated sample DVH's and report dosimetric parameters to offer clinicians a unique dosimetric perspective. (c) 2019 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.
机译:乳房术中放射治疗(IORT)与IntraBeam系统使用50kV X射线源,在哺乳期手术期间将单一的放射疗法施加到肿块切除术腔中。我们寻求对刚性球形涂抹器的乳胶切除术腔进行剂量分析。采用水幻影测量来验证供应商提供的X射线校准。规划目标体积(PTV)定义为球形涂敷器的10mm膨胀,产生剂量直方图(DVH)和剂量参数[D-min,D-1mm,V-90,V-90报道,V-50,HI]报道。另外,使用90%和80%同学水平的治疗方法[R-90,R-80]。当深度剂量(PDD)归零时,较小的涂敷器具有陡峭的PDD。对于涂抹器表面的处方剂量,PTV的剂量参数范围为:3.15至6.84GY,D-MIN为D-1mm,比-1MM为16.2至17.6GY,V为2.6至6.9% V-80的-90,5.5至15.1%,V-50的21.1至55.6%。对于直径15至50mm的涂抹器,HI,R-90的HI,0.53至0.85mm为0.53至0.85mm,R-80为1.18至1.85mm。较小的涂抹器具有降低的PTV覆盖率,但升高,因为近端源的光束的衰减更加明显。另外,小涂抹器(<= 30mm)的铝过滤器的存在增加了PTV覆盖率,但降低了涂抹器表面上的剂量率。 IORT的交付是在或不使用基于图像的规划的情况下进行的。为了克服这一限制,我们已经生成了样本DVH并报告了Dosimetric参数,以提供临床医生独特的DoSimetric视角。 (c)2019年美国医学剂量分子协会。由elsevier Inc.保留所有权利发布。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号