Purpose To develop a robust and flexible low power water excitation pulse that enables effective fat suppression at high magnetic field strength. Methods A water excitation method that uses spatially non‐selective pulses was optimized in numerical simulations, and implemented and tested in phantoms and healthy volunteers at 3T. The lipid insensitive binomial off‐resonant excitation (LIBRE) pulse comprises two low power rectangular subpulses that have a variable frequency offset, phase offset and duration. The capability and extent of LIBRE fat suppression was quantitatively compared with conventional fat saturation (FS) and water excitation (WE) techniques. Results LIBRE enables simultaneous water excitation and near complete fat suppression in large volumes at 3T as demonstrated by numerical simulations, and experiments. In phantoms and in human subjects, the frequency responses matched well with those from the numerical simulation. Comparing FS and WE, LIBRE demonstrated an improved robustness to magnetic field inhomogeneities, and a much more effectively suppressed fat signal. This applied for a range of pulse durations and pulses as short as 1.4 ms. Conclusion A flexible water excitation method was developed that shows robust, near complete fat suppression at 3T. Magn Reson Med 79:3007–3017, 2018. ? 2017 International Society for Magnetic Resonance in Medicine.
机译:目的,开发一种坚固且灵活的低功率水励磁脉冲,可实现高磁场强度的有效脂肪抑制。方法使用空间非选择性脉冲在数值模拟中优化使用空间非选择性脉冲的水激发方法,并在3T的幻影和健康志愿者中实施和测试。脂质不敏感的二项式偏离谐振激发(Libre)脉冲包括两个低功率矩形子样脉冲,其具有可变频率偏移,相位偏移和持续时间。与常规脂肪饱和度(FS)和水激发(我们)技术相比,自由脂肪抑制的能力和程度。结果Libre在3T时使同时水激发和在大量的大量抑制附近,如数值模拟所示,以及实验。在幽灵和人类受试者中,频率响应与数值模拟的频率匹配良好。比较FS和我们,Libre展示了对磁场不均匀性的改进的鲁棒性,并且更有效地抑制了脂肪信号。这施加了一系列脉冲持续时间和脉冲,短至1.4毫秒。结论开发了一种柔性水励磁法,其表现出稳健,近3T完全脂肪抑制。 Magn Reson Med 79:3007-3017,2018.? 2017年医学磁共振的国际社会。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号