首页> 外文期刊>Graefe's archive for clinical and experimental ophthalmology: Albrecht von Graefes Archiv fur klinische und experimentelle Opthalmologie >Misleading axial length measurements with the IOLMaster due to a dense posterior vitreous surface membrane and a macular edema in a diabetic patient
【24h】

Misleading axial length measurements with the IOLMaster due to a dense posterior vitreous surface membrane and a macular edema in a diabetic patient

机译:由于糖尿病患者的致密玻璃体表面膜和黄斑水肿而误导轴向长度测量。在糖尿病患者中的黄斑水肿

获取原文
获取原文并翻译 | 示例

摘要

Exact biometry is an essential component in cataract surgery to obtain accurate refractive results [1]. The IOLMaster? is a precise and common instrument to obtain axial length [2-5]. We describe a case of a patient with diabetic retinopathy in which the IOLMaster? was not able to determine an axial length. Clear signals from a dense posterior membrane as well as from retinal pigment epithelium (RPE) and macular edema were received.
机译:精确的生物测定是白内障手术中的必要组分,以获得准确的折射率[1]。 iolmaster? 是获得轴向长度[2-5]的精确和常见的仪器。 我们描述了患有糖尿病视网膜病的病例,其中IOLMASTER? 无法确定轴向长度。 接收来自致密后膜的清除信号以及从视网膜颜料上皮(RPE)和黄斑水肿。