Comparison of Goldmann applanation and Ocular Response Analyser tonometry: intraocular pressure agreement and patient preference

机译:Goldmann Applanation和眼响应分析仪Tononetry的比较:眼内压力协议与患者偏好

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives To evaluate the agreement between Goldmann applanation tonometry (GAT) and Ocular Response Analyser (ORA) intraocular pressure (IOP) measurements, and patients' preferences. Methods Both eyes of participants in the 'Glaucoma within the Northern Ireland Cohort for the Longitudinal Study of Ageing' (GwNICOLA) were included. Participants underwent GAT by a glaucoma expert and ORA tonometry in a random order. Investigators were masked to measurements between devices. Participants were asked which tonometer, if any, they would prefer. We estimated the 95% limits of agreement (95% LoA) and the variables that influence agreement between tonometers. Results There were 228 eyes of 120 participants included in this study. Mean age of participants was 68.0 years (SD 8.79) and 52.5% were female. For GAT-ORA IOPcc the mean difference with GAT (95% CI) was -0.23 mmHg (-0.57 mmHg, 0.11 mmHg) and the 95% LoA (95% CIs) were from 4.82 mmHg (5.15 mmHg, 4.48 mmHg) to -5.28 mmHg (-5.61 mmHg, -4.94 mmHg). 40.8% of eyes had an IOP difference of 2 mmHg or more between GAT and ORA IOPcc. Corneal resistance factor (CRF) as estimated by ORA influenced the agreement between GAT and ORA IOPcc. There were no differences in preference for method of tonometry. Conclusions Although ORA IOPcc measurements with ORA did not show significant bias compared with GAT, the relatively large proportion of measurement differences between ORA IOPcc and GAT that were >2 mmHg indicates that GAT and ORA IOP measurements may not be interchangeable. There were no differences in preference for method of tonometry.
机译:目标旨在评估金曼施静脉曲线(GAT)和眼响应分析仪(OOP)测量和患者偏好的协议。方法包括核查北爱尔兰北部青光眼的参与者眼睛,致老化纵向研究(Gwngola)。参与者通过青光眼专家和ora Tononetry以随机顺序接受GAT。调查人员被掩盖到设备之间的测量。被问及参与者,如果有的话,他们更愿意。我们估计了95%的协议限制(95%LOA)和影响TONOMENTER之间协议的变量。结果本研究中包含120名参与者的228只眼睛。参与者的平均年龄为68.0岁(SD 8.79)和52.5%是女性。对于GAT-OOPCC与GAT(95%CI)的平均差异为-0.23mmHg(-0.57mmHg,0.11mmHg),95%LOA(95%CIS)为4.82mmHg(5.15mmHg,4.48mmHg) - 5.28 mmhg(-5.61 mmhg,-4.94 mmhg)。 40.8%的眼睛在GAT和OOPCC之间具有2mmHg或更多的IOP差异。 ORA估计的角膜抵抗因子(CRF)影响了GAT和OOPCC之间的协议。偏好对纯度的方法没有差异。结论虽然与GAT相比,OOPCC测量与OOPCC测量没有显示出显着的偏差,但2mmHg的OOPCC和GAT之间的测量差异相对大的测量差异表明GAT和OOP测量可能不可互换。偏好对纯度的方法没有差异。

著录项

  • 来源
    《Eye》 |2020年第3期|共7页
  • 作者

  • 作者单位
  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 眼科学;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号