首页> 外文期刊>Scientific reports. >Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients
【24h】

Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients

机译:患者测量的眼压协议使用回弹正文与Golaucoma Applanation Tononetry(Gat)在青光眼患者中的反弹

获取原文
           

摘要

This study aims to determine the agreement of patient-measured intraocular pressure (IOP) using rebound tonometry with ophthalmologist-measured IOP using Goldmann applanation tonometry (GAT). Fifty-three glaucoma patients used rebound tonometry (Icare ONE, Icare Finland Oy., Finland) to measure their own IOP in ambient environments for 1 week, 5 times per day. Clinic IOP measurements were performed by ophthalmologists using GAT and by patients using rebound tonometry on examination days 1, 4 and 7 of the same week. The agreement between the two tonometries was evaluated by modified Bland-Altman plots and intra-class correlation coefficient (ICC) was determined. Differences in ICCs of them among the three examination days were evaluated by bootstrap resampling analysis. Respective within-measurement ICC of GAT and rebound tonometry were 0.98 and 0.94 on Day 1, 0.98 and 0.93 on Day 4, and 0.96 and 0.92 on Day 7. In a modified Bland-Altman plot, the mean difference ±1 standard deviation (SD) between the two tonometries was 0.15?±?0.65?mmHg (p?=?0.682). Between-measurement ICC were 0.66, 0.76 and 0.73 on the 3 examination days. There was no significant difference among ICCs. In conclusion, patient-measured IOP using rebound tonometry and ophthalmologist-measured IOP using GAT demonstrate good agreement.
机译:本研究旨在使用金属师施加型(GAT)来确定使用具有眼科医生测量的眼压仪测量的患者测量的眼压(IOP)的协议。五十三名青光眼患者使用反弹纯度(ICare One,Icare Finland Oy。,芬兰)在环境环境中测量自己的IOP 1周,每天5次。临床IOP测量由眼科医生使用GAT和在同一周的考试时第1,4和7日使用反弹型曲目的患者进行。通过改进的平坦-Attman图和阶级相关系数(ICC)评估了两种曲纹之间的协议。通过引导重采样分析评估三个检查日中ICC的差异。在第1天,第4天,第7天,第7天,第7天,第7天,0.98和0.93时,0.98和0.93的测量内的测量内的ICC各自为0.98和0.94.在修改的平坦 - 奥特曼图中,平均差异±1标准偏差(SD )两种曲纹之间为0.15?±0.65?mmHg(p?= 0.682)。在3个检查日,测量ICC为0.66,0.76和0.73。 ICCS之间没有显着差异。总之,使用GAT使用回弹纯度和眼科医生测量的IOP测量的IOP表现出良好的一致性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号