IntroductionExternal fixation is associated with the risk of pin loosening and pin infection potentially associated to thermal bone necrosis during pin insertion.ObjectiveThis study aims to investigate if the use of external fixator systems with unicortical pins reduces the heat production during pin insertion compared to fixators with bicortical pins.MethodsPorcine bone specimens were employed to determine bone temperatures during insertion of fixator pins. Two thermographic cameras were used for a simultaneous temperature measurement on the bone surface (top view) and a bone cross-section (front view). Self-drilling unicortical and bicortical pins were inserted at different rotational speeds: (30-600) rpm. Maximum and mean temperatures of the emerging bone debris, bone surface and bone cross-section were analyzed.ResultsMaximum temperatures of up to 7726 degrees C were measured during pin insertion in the emerging debris and up to 42 +/- 2 degrees C on the bone surface. Temperatures of the emerging debris increased with increasing rotational speeds. Bicortical pin insertion generated significantly higher temperatures at low insertion speed (30rpm)ConclusionThe insertion of external fixator pins can generate a considerable amount of heat around the pins, primarily emerging from bone debris and at higher insertion speeds. Our findings suggest that unicortical, self-drilling fixator pins have a decreased risk for thermal damage, both to the surrounding tissue and to the bone itself.
机译:销售外部固定与引脚松动和引脚感染有关的销钉感染有关,销钉插入过程中可能与热骨坏死有关。目的研究旨在调查外部固定器系统是否使用单髓销期间的销插入期间的热量产生,与具有双模式的固定器相比销插入期间的热量生产PINS.Methodsporcine骨标本用于在插入固定器销期间确定骨温度。两个热选摄像机用于骨表面(顶视图)和骨横截面(前视图)上的同时温度测量。以不同的旋转速度插入自钻Unicorcical和Bicortical销:(30-600)RPM。分析了新出现的骨碎片,骨表面和骨横截面的最大和平均温度。在新出现的碎片销插入期间测量高达7726摄氏度的高达7726摄氏度,骨骼上最多42 +/- 2℃表面。新兴碎片的温度随着旋转速度的增加而增加。 Bicortical销插入在低插入速度(30rpm)结束时产生明显较高的温度(30rpm)结论外固定器销的插入可以在销周围产生相当大量的热量,主要从骨碎片和更高的插入速度出现。我们的研究结果表明,Unicormical钻孔固定器销具有降低的热损伤的风险,无论是周围组织和骨骼本身。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号