Quantification of enhancement in contrast-enhanced spectral mammography using a custom-made quantifier tool ( I -STRIP): A proof-of-concept study

机译:使用定制量化型工具(i-strip)的对比度增强光谱乳房X线照相X线图的量化量化:概念证明研究

获取原文
获取原文并翻译 | 示例
           

摘要

PurposeContrast-enhanced spectral mammography (CESM) is diagnostically superior to full-field digital mammography. An important improvement for CESM would be the ability to quantify enhancement. In this proof-of-concept study we present a method for quantifying CESM enhancement. Methods and materialsWe developed a custom-made quantifier tool (I-STRIP) containing five chambers, each filled with a different iodine mass thickness (IMT). CESM grey values of the recombined image (CGV) in theI-STRIP were used to quantify breast IMT. We evaluated theI-STRIP’s accuracy using a dedicated breast phantom containing chambers with known IMT’s. Furthermore, we tested the effect of theI-STRIP on image quality and clinical use in five patients. Retrospectively, we studied whether current CESM protocols could distinguish between malignant and benign lesions in terms of CGV. ResultsPhantom experiments showed that quantification was independent of chamber height and size, phantom thickness andI-STRIP position for different IMT’s (1.5, 3.0 and 7.5?mg l/cm2). Near the phantom’s periphery accuracy was found to be lower due to the breast-within-breast artifact. In the clinical setting (n?=?5), workflow and image quality were not influenced by theI-STRIP. The mean IMT of these invasive breast cancers was 2.1?mg l/cm2(range 1.3–3.4?mg?l/cm2). Malignant lesions showed significantly higher CGV’s than benign lesions (p?=?0.002). ConclusionWe showed in both phantom and clinical experiments that CESM quantification is feasible, without influencing workflow or image quality. The current CESM imaging protocol seems to be able to distinguish between benign and malignant breast lesions in terms of CGV.
机译:PurposeContrast-Enhanced光谱乳房X线照片(CESM)是诊断上优于全场数字乳房X线摄影。 CESM的重要改进将是量化增强的能力。在该概念证明研究中,我们提出了一种定量CESM增强的方法。方法和材料开发了一种包含五个腔室的定制定量型工具(I-Stab),每个腔室填充有不同的碘质量厚度(IMT)。 TheI-带中的重组图像(CGV)的CESM灰度值用于量化乳腺IMT。我们评估了使用具有已知IMT的专用乳房幻影的专用乳房幻影的精度。此外,我们在五名患者中测试了Thei-Strip对图像质量和临床应用的影响。回顾性地,研究当前CESM协议是否可以区分恶性和良性病变。结果露出的实验表明,定量与腔室高度和尺寸无关,幻象厚度和不同IMT的位置(1.5,3.0和7.5?Mg L / cm2)。由于乳房内术文物,发现近距离幻像的外围准确性降低。在临床环境中(n?=?5),工作流程和图像质量不受The-Strip的影响。这些侵入性乳腺癌的平均IMT为2.1?mg L / cm 2(范围1.3-3.4Ω·mg?l / cm 2)。恶性病变显示出明显高于良性病变(P?= 0.002)。结论我们在幻影和临床实验中显示,CESM量化是可行的,而不会影响工作流程或图像质量。目前的CESM成像协议似乎能够在CGV方面区分良性和恶性乳腺病变。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号