首页> 外文期刊>Echocardiography. >The first case of “locked‐in leaflet” after transcatheter aortic valve replacement in a patient with bicuspid aortic stenosis
【24h】

The first case of “locked‐in leaflet” after transcatheter aortic valve replacement in a patient with bicuspid aortic stenosis

机译:经触发表主动脉瓣膜置换患者的第一种“锁定叶片”的第一种案例,患有双囊主动脉狭窄

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

This case report is about an 85‐year‐old woman with bicuspid aortic stenosis ( AS ). Although preoperative multimodality imaging showed challenging anatomical aspects, transcatheter aortic valve replacement ( TAVR ) was selected to be performed as a less invasive alternative treatment approach, owing to her comorbidity. Postoperative transesophageal echocardiography and multidetector‐row computed tomography revealed the presence of “locked‐in leaflet” caused by stent distortion due to pinching by calcified native leaflets, with nodular calcification preventing the full expansion of the valve, which resulted in moderate perivalvular leakage. This is the first reported case of bicuspid AS treated with TAVR that eventually resulted in “locked‐in leaflet.”
机译:本案例报告是关于一名85岁的女性,具有双裂主动脉狭窄(AS)。 尽管术前多模成像显示出具有挑战性的解剖学方面,但由于她的合并症,选择了经转丝管主动脉瓣更换(TAVR)作为较少的侵入性替代治疗方法进行。 术后经细胞深呼超声心动图和多传输线计算断层摄影揭示了由于通过钙化的天然传单夹持的支架失真引起的“锁定叶片”的存在,具有结节钙化,防止了阀的全膨胀,从而导致中度恐吓泄漏。 这是第一个报告的Bicuspid案例,如Tavr处理,最终导致“锁定传单”。