首页> 外文期刊>Investigative radiology >Accuracy and Repeatability of Automated Injector Versus Manual Administration of an MRI Contrast Agent—Results of a Laboratory Study
【24h】

Accuracy and Repeatability of Automated Injector Versus Manual Administration of an MRI Contrast Agent—Results of a Laboratory Study

机译:自动注射器的准确性和可重复性与手动管理MRI造影剂 - 实验室研究结果

获取原文
获取原文并翻译 | 示例

摘要

Objective: The aim of this study was to compare flow rates over time and the deviations from the target flow rate of a magnetic resonance imaging contrast agent achieved by an automated injector versus manual injection.Materials and Methods: In this laboratory study, the magnetic resonance contrast agent gadobutrol was repeatedly injected by an injector and by 10 experienced technologists. Six scenarios with 2 different target flow rates (1 and 5 mL/s), 2 different contrast volumes (10 and 20 mL), and 2 different intravenous (IV) catheters (22 gauge and 20 gauge) were tested. The flow rates over time were recorded. The target variable was the average absolute deviation and average absolute percentage deviation from the target flow rate.Results: The flow rates over time achieved by an injector were almost identical. Slight deviations from the target flow rate occurred during ramp-up and ramp-lown only. Those of manual injection showed high variability over the whole course of the injection. In the 1 mL/s scenarios, the injector deviated from the target flow rate by 0.06 mL/s or less (≤6%) and in the 5 mL/s scenarios 1.02 mL/s or less (<20%). For the manual injections at the same flow rates, these figures were 0.35 mL/s or less (≤35%) and 3.1 mL/s or less (≤62%).Conclusions: Injector administration of a magnetic resonance contrast agent ninimally deviated from the target flow rate, whereas manual injection varied widely. Injector administration is more accurate and repeatable.
机译:目的:本研究的目的是将流速随时间的比较和偏离自动喷射器与手动注射的磁共振成像造影剂与手动注射的磁共振成像造影剂的偏差。材料和方法:在该实验室研究中,磁共振用喷射器反复注入造影剂GadoButrol,并通过10位经验丰富的技术人员重复注入。具有2个不同的目标流量速率(1和5 mL / s),2种不同对比度(10和20mL)的六种场景,以及2个不同的静脉内(IV)导管(22号和20尺)。记录随时间的流速。目标变量是与目标流量的平均绝对偏差和平均绝对百分比偏差。结果:喷射器实现的随时间随时间的流速几乎相同。从斜坡和斜坡的斜坡期间发生略有偏差。手动注射的那些在整个注射过程中显示出高度的变化。在1mL / s情景中,喷射器偏离目标流速达0.06ml / s或更低(≤6%),在5 ml / s场景中偏离1.02ml / s以下(<20%)。对于相同的流量速率的手动注射,这些图为0.35ml / s或更低(≤35%)和3.1ml / s或更低(≤62%)。结论:喷射器施用磁共振造影剂九十偏离目标流量,而手动注射差异很大。喷射器给药更准确和可重复。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号