Comparison of performance and abnormal cell flagging of two automated hematology analyzers: Sysmex XN XN 3000 and Beckman Coulter DxH 800

机译:两种自动血液学分析仪性能和异常细胞标记的比较:Sysmex XN XN 3000和Beckman Coulter DXH 800

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Introduction The purpose was to evaluate the analytical performances of Sysmex XN ?3000 and UniCel DxH 800 comparing the obtained results with manual counting and between each other. Also flagging capabilities of abnormal cells were compared for both analyzers. Methods Two thousand one hundred and forty‐two whole‐blood samples were analyzed for evaluation. The samples flagged due to blast, atypical lymphocyte ( AL ), immature granulocyte ( IG ), or nucleated red blood cells ( NRBC ) were microscopically reviewed (n=102). Results The within‐run and between‐day coefficient of variations ( CV %) of XN 3000 for hemoglobin, RBC , MCV , WBC , and platelets were lower than 5% and for WBC differentials lower than 10% except monocytes which was 15.6% at low level. The precision results of DxH 800 were also lower than 5.0% except platelets (9.5%) and monocytes (45%) at low level. The comparison of analyzers revealed good agreement ( R .86), except monocytes and basophils. The flagging sensitivities of XN 3000 were higher for IG s, blasts, and AL s than those of DxH 800 and almost similar for NRBC . Conclusion The XN 3000 and DxH 800 are accurate, highly precise systems and can be used effectively in high‐volume laboratories. The flagging sensitivity of XN 3000 was higher in detecting blasts, IG s, and AL s than that of DxH 800. The detection of abnormal cells with high sensitivity may improve laboratory workflow with a reduced slide review and accelerated turnaround time.
机译:摘要介绍该目的是评估Sysmex XN的分析性能,Unicel DXH 800比较所获得的结果与手动计数和彼此之间的比较。对两种分析仪进行了比较了异常细胞的标记能力。方法分析了两千一百四十二个全血样品进行评价。由于Blast,非典型淋巴细胞(Al),未成熟的粒细胞(Ig)或核细胞红细胞(NRBC)标记的样品是显微诊断的(n = 102)。结果血红蛋白,RBC,MCV,WBC和血小板的XN 3000的XN 3000内的XN 3000之间的运行情况和白天的变化系数(CV%)低于5%,除了单核细胞之外,对于15.6%的单核细胞,WBC差异低于10%低级。除血小板(9.5%)和单核细胞(45%)下,DXH 800的精度结果也低于5.0%,低水平。分析仪的比较显示出良好的协议(R>除了单核细胞和嗜碱性粒细胞外。对于IGS,Blasts和Al S的XN 3000的标记敏感性比DXH 800的标记敏感性更高,并且对于NRBC几乎相似。结论XN 3000和DXH 800是精确的,高精度的系统,可有效地用于大批量实验室。检测爆炸,IG S和Al S的XN 3000的标记敏感性高于DXH 800的敏感性。具有高灵敏度的异常细胞的检测可以改善实验室工作流程,减少幻灯片审查和加速周转时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号