Acquired Von Willebrand Syndrome and Blood Pump Design

机译:获得了冯维尔布兰综合征和血泵设计

获取原文
获取原文并翻译 | 示例
       

摘要

The existence of acquired von Willebrand syndrome (AVWS) in patients with continuous flow left ventricular assist devices (LVADs) is well documented and has been verified by numerous investigators. AVWS has not been observed to occur in pulsatile devices such as the SynCardia total artificial heart (TAH), the HeartMate XVE, and the Thoratec pulsatile ventricular assist device (PVAD) used as a single pump. AVWS can also occur in patients with aortic stenosis, ventricular septal defect, mitral stenosis, and patent ductus arteriosus. It has been experimentally verified that supraphysiologic shear stress that occurs under these conditions can cleave the von Willebrand molecule, but the critical magnitude of stress and duration is unclear. Limited experimental results demonstrate that shear stresses as low as 5 Pa (50 dyne/cm(2)) can cause cleavage. Stresses in current centrifugal pumps can be as high as two orders of magnitude greater than this value. Pulsatile LVADs have stresses almost two orders of magnitude less than continuous flow LVADs. In order to improve continuous flow LVADs, the challenge for designers is to first determine the magnitude and duration of stress that is causing AVWS and then, if possible, design a pump below these stresses.
机译:患有连续流动左心室辅助装置(LVADS)患者的获得的von Willebrand综合征(AVWS)的存在是充分记录的,并通过许多调查人员验证。在脉动装置中未观察到诸如Syncardia总人造心脏(TAH),骨髓XVE和肌炎脉动室辅助装置(PVAD)的脉冲装置中未被观察到。 AVWS还可以在主动脉狭窄,心室隔膜缺损,二尖瓣狭窄和专利导管患者中进行。已经通过实验验证,在这些条件下发生的超级剪切应力可以切割von willebrand分子,但应力和持续时间的临界幅度尚不清楚。有限的实验结果表明,低至5Pa(50达因/厘米(2))的剪切应力会导致切割。电流离心泵的应力可以高达两个大小大于该值的数量级。 Pulsatile LVADS强调几乎比连续流动LVAD的数量数几乎两个数量级。为了改善连续流量的LVAD,设计师的挑战是首先确定导致AVWS的应力的幅度和持续时间,然后在可能的情况下设计泵以下。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号