AbstractA microfluidic paper-based analytical device (μPADs) immunoassay for detection of the blood native biomarker D-dimer is reported. The μPAD is created by wax printing on a single piece of chromatographic paper and combined with an anti-D-dimer capture antibody and conjugates of anti-D-dimer antibody with 40?nm gold nanoparticles. The presence of D-dimer in buffer/simulated plasma samples is successfully reported for concentrations as low as 15?ng D-dimer/mL. Linearity between signal intensity and D-dimer concentration is observed up to 100?ng/mL. Using an appropriate dilution, the test could be used to yield positive results only for those samples with a D-dimer concentration above the clinically relevant threshold range of 250–500?ng/mL. Finally, the merits and pitfalls of using μPADs as compared to lateral flow devices in immunoassays are discussed.]]>
机译:<![CDATA [ 抽象 报告了用于检测血液天然生物标志物D-二聚体的微流体纸的分析装置(μpads)免疫测定。通过蜡印刷在单个色谱纸上产生μPAD,并与抗D-二聚体抗体和40·NM金纳米颗粒的抗D-二聚体抗体的缀合物结合。成功地将D-二聚体在缓冲液/模拟等离子体样品中的存在以低至15〜Ng D-二聚体/ mL的浓度。信号强度和D-二聚体浓度之间的线性程度可观察到100μg/ ml。使用适当的稀释度,测试可用于仅对那些具有D-二聚体浓度的样品产生阳性结果,以上临床相关的阈值范围为250-500μg/ mL。最后,讨论了与免疫测定中的横向流动装置相比使用μPAD的优点和缺陷。 ]]>