Relative accuracy of computerized intrapartum fetal heart rate pattern recognition by ultrasound and abdominal electrocardiogram detection

机译:超声和腹心电图检测计算机电脑胎儿心率模式识别的相对准确性

获取原文
获取原文并翻译 | 示例
           

摘要

Abstract Introduction Noninvasive fetal heart rate monitoring using transabdominal fetal electrocardiographic detection is now commercially available and has been demonstrated to be an effective alternative to traditional Doppler ultrasonographic techniques. Our objective in this study was to compare the results of computerized identification of fetal heart rate patterns generated by ultrasound‐based and transabdominal fetal electrocardiogram‐based techniques with simultaneously obtained fetal scalp electrode‐derived heart rate information . Material and methods We applied an objective computer‐based analysis for recognition of fetal heart rate patterns (Monica Decision Support) to data obtained simultaneously from a direct fetal scalp electrode, Doppler ultrasound, and the abdominal‐fetal electrocardiogram techniques. This allowed us to compare over 145?hours of fetal heart rate patterns generated by the external devices with those derived from the scalp electrode in 30 term singleton uncomplicated pregnancies during labor. The direct fetal scalp electrode is considered to be the most accurate and reliable technique used in current clinical practice, and was, therefore, used as the standard for comparison. The program quantified the baseline heart rate, long‐ and short‐term variability. It indicated when an acceleration or deceleration was present and whether it was large or small. Results Ultrasound was associated with significantly greater deviations from the fetal scalp electrode results than the abdominal fetal electrocardiogram technique in recognizing the correct baseline heart rate, its variability, and the presence of small and large accelerations and small decelerations. For large decelerations the two external methods were each not significantly different from the scalp electrode results. Conclusions Noninvasive fetal heart rate monitoring using maternal abdominal wall electrodes to detect fetal cardiac activity more reliably reproduced the computerized analysis of heart rate patterns derived from a direct fetal scalp electrode than did traditional ultrasound‐based monitoring. Abdominal‐fetal electrocardiogram should, therefore, be considered a primary option for externally monitored patients.
机译:摘要介绍使用Transabominal胎儿心电图检测的非侵入性胎儿心率监测现在可商购获得,并且已被证明是传统多普勒超声技术的有效替代方案。我们在本研究中的目的是比较通过同时获得基于胎儿头皮电极衍生的心率信息的超声基和转基因胎儿心电图技术产生的胎儿心率模式的计算机化鉴定结果。材料和方法我们应用了基于电脑的分析,以识别胎儿心率模式(MONICA决策支持)与直接胎儿头皮电极,多普勒超声波和腹部心电图技术同时获得的数据。这使我们可以比较由外部装置产生的超过145小时的胎儿心率模式,其中包含在劳动期间的30次术语单一简单的妊娠中源于头皮电极的那些。直接胎儿头皮电极被认为是当前临床实践中最准确和可靠的技术,因此用作比较标准。该计划量化了基线心率,长期和短期变异性。它在存在加速或减速时表明,是否大或小。结果超声与胎儿头皮电极的偏差明显更大,而不是腹部心电图技术认识到正确的基线心率,其可变性和小型加速度和小型减速的存在。对于大型减速,两个外部方法与头皮电极结果无显着不同。结论使用母体腹壁电极检测胎儿心脏活性的非侵入性胎儿心率监测更可靠地再现从直接胎儿头皮电极的心率模式的计算机化分析而不是传统的超声波的监测。因此,腹部心电图应被视为外部监测患者的主要选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号