Comparison of Flow Cytometric Methods for the Enumeration of Residual Leucocytes in Leucoreduced Blood Products: A Multicenter Study

机译:幼苗血液产物中残留白细胞枚举流动细胞术方法的比较:多中心研究

获取原文
获取原文并翻译 | 示例
           

摘要

The BD FACSVia (TM) System features novel designs in hardware, software, and instrument QC. We compared the performance of the BD FACSVia System using the BD Leucocount (TM) kit with the BD FACSCalibur (TM) flow cytometer. Leucoreduced platelet (PLT, n = 252) and red blood cell (RBC, n = 278) specimens were enrolled at four sites. Each specimen was stained in four tubes using the BD Leucocount kit reagents and acquired on the two systems. BD Leucocount Control cells (high and low) were used to evaluate the inter-site reproducibility on the BD FACSVia System at three sites over 20 days. Deming regression and Bland-Altman analysis were performed to determine the WBC absolute counts on the BD FACSVia System vs. the BD FACSCalibur system. Assay accuracy for the range of 0-350 WBCs/mu l was adequate. For samples with 25 WBCs/mu l, the bias with 95% limits of agreement was 0.136 (-1.897 to 2.169) WBC/mu l for PLTs (n = 184) and 0.170 (-2.025 to 2.365) WBC/mu l for RBCs (n = 193). For inter-site reproducibility, the CV% was 6.46% (upper 95% CI 7.16%) for the PLT high control and 9.49% (10.52%) for the PLT low control. The CV% was 7.51% (8.32%) for the RBC high control and 10.76% (11.92%) for the RBC low control. The BD FACSVia System reported equivalent results of WBC absolute counts for leucoreduced PLT and RBC samples compared to the BD FACSCalibur system. The inter-laboratory reproducibility of the BD FACSVia System met study specifications. (C) 2018 The Authors. Cytometry Part A Published by Wiley Periodicals, Inc. on behalf of ISAC.
机译:BD Facsvia(TM)系统具有新颖的硬件,软件和仪器QC设计。我们将BD Facsvia系统使用BD Leucocount(TM)套件与BD FacScalibur(TM)流式细胞仪进行了比较。在四个位点注册了睫状血小板(PLT,N = 252)和红细胞(RBC,N = 278)样品。每个样品使用BD Leucocount试剂试剂在四管中染色并在两种系统上获得。 BD Leucocount控制细胞(高低)用于评估在20天内的三个地点的BD Facsvia系统的现场间再现性。进行了脱位回归和Bland-Altman分析,以确定BD Facsvia系统对BD FacScibur系统的WBC绝对计数。测定精度为0-350 WBCS / MU L充足。对于具有&的样品。 25 WBCS / MU L,具有95%的协议限制的偏差为0.136(-1.897至2.169)WBC / mu L用于PLTS(n = 184)和0.170(-2.025至2.365)WBC / MU L对于RBCS(n = 193)。对于位分区的再现性,CV%为PLT高对照的6.46%(上95%CI 7.16%),PLT低对照为9.49%(10.52%)。 CV%为RBC高对照的7.51%(8.32%),RBC低对照的10.76%(11.92%)。与BD FACSCALIBUR系统相比,BD Facsvia系统报道了白细胞传导的PLT和RBC样品的WBC绝对计数的等效结果。 BD Facsvia系统的实验室再现性符合研究规范。 (c)2018作者。 Cytometry部分由Wiley期刊,Inc。代表ISAC发布。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号