Abstract Background The purpose of this study was to use a physiological pressure transducer to measure real‐time, continuous pressure changes in an ex vivo study model of porcine eyes to record the amount of force needed for scleral penetration and to measure acute intraocular pressure rise during intravitreal injections. Methods A pressure transducer was inserted into the anterior chamber of 30 fresh porcine eyes, and intraocular pressure was measured 2?s prior to intravitreal injection until 2?s after. A force transducer plate was used to insert various gauge needles into the vitreous cavity and the amount of force in Newtons (N) required for scleral penetration was recorded. Results For scleral perforation, 32‐ and 30‐gauge needles required 0.44?N and 0.45?N, significantly less than larger gauge needles ( P ??0.05). Similarly, 27‐ and 25‐gauge needles required more force than smaller gauge needles but less than 19 gauge ( P ??0.05). Intraocular pressure increased an average of 64.5?mmHg during intravitreal injection. Two seconds postinjection intraocular pressure readings showed a residual intraocular pressure increase of 11.1?mmHg from pre‐injection baseline. Conclusion Real‐time continuous recordings of pressure reveal that an instantaneous intraocular pressure spike occurs during intravitreal injection and appears to be separate from the intraocular pressure spike that occurs during needle insertion. This pressure spike is transient and has not been captured by previous methods of intraocular pressure measurement, which rely on single time point measurements. The clinical significance of this brief intraocular pressure spike is unclear and warrants further investigation.
机译:摘要背景本研究的目的是使用生理压力传感器来测量猪眼的前体内研究模型的实时,连续压力变化,以记录巩膜渗透所需的力量,并测量急性眼内压力玻璃体内注射。方法将压力传感器插入30个新鲜猪眼的前室中,并在玻璃体玻璃体注射之前测量2℃的眼内压直至2°。使用力传感器板将各种规格针插入玻璃体中,记录巩膜渗透所需的牛顿(n)中的力量。巩膜穿孔的结果,需要32-和30·仪表针,需要0.44Ω·n和0.45?n,显着小于较大的仪表针(p≤≤0.05)。类似地,27-和25·仪表针比较小的仪表针更加力量,但小于19尺(p≤≤0.05)。在玻璃体内注射期间,眼内压平均增加64.5ΩmmHg。两秒钟关注的眼压读数显示出从预注射基线的剩余眼压增加11.1Ωmmhg。结论压力的实时连续记录表明,在玻璃体内注射期间发生瞬时眼压尖峰,似乎与针插入期间发生的眼压尖峰分开。这种压力尖峰是瞬态的,并且尚未被先前的眼内压测量方法捕获,这依赖于单时间点测量。这种短暂的眼内压穗的临床意义尚不清楚并保证进一步调查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号