首页> 外文期刊>Acta orthopaedica Scandinavica. >Accuracy of radiographic and radiostereometric wear measurement of different hip prostheses: an experimental study.
【24h】

Accuracy of radiographic and radiostereometric wear measurement of different hip prostheses: an experimental study.

机译:不同髋关节假体的X线和X线立体测量磨损测量的准确性:一项实验研究。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: In vivo measurement of wear in the ball and socket articulation of total hip arthroplasties is of interest in the evaluation of both existing and new implants. Controversy reigns regarding the accuracy of different radiological measurement techniques and in particular how accuracy has been assessed. MATERIAL AND METHODS: We assessed the accuracy of 2 radiostereometric (RSA) techniques for wear measurement and 3 standard radiographic techniques, namely Imagika (image analyzing software), Imagika corrected for head center displacement, and the Charnley Duo method. 5 custom-made adjustable phantoms with different prosthetic components were used. RESULTS: In 20 measurements of all 5 phantoms at 3 levels of simulated wear (0.2 mm, 1.0 mm and 1.5 mm), the mean measurement error of the digital RSA examinations was 0.010 mm (accuracy 0.42). The corresponding error values for the three radiographic techniques were 0.19 (accuracy 1.3) for Charnley Duo, 0.13 (accuracy 1.3) for Imagika corrected, and 1.021 (accuracy 2.99) for Imagika. Measurement error decreased from 0.011 mm with ordinary RSA to 0.004 with RSA digital measurement. Head size, direction of wear in relation to the cup or type of prosthetic component did not influence the measurement error. The results of Charnley Duo and Imagika corrected were similar but the latter had an inexplicable systematic error in measuring one of the phantoms. Imagika had the worst results due to its inability to compensate for the out-of-head center effect. Alumina heads were difficult to analyze with all methods. INTERPRETATION: By using the ISO standard for assessing accuracy, RSA can be expected to measure wear with an accuracy of about 0.4 mm irrespective of prosthetic component studied or direction of wear, whereas the best technique, in our study, based on standard radiographs can be accurate to about 1.3 mm.
机译:背景:全髋关节置换术的球窝关节关节内磨损的体内测量在评估现有植入物和新型植入物方面都令人关注。关于不同的放射线测量技术的准确性,尤其是如何评估准确性的争议颇多。材料与方法:我们评估了2种用于测量磨损的放射线立体(RSA)技术和3种标准射线照相技术的准确性,即Imagika(图像分析软件),针对头部中心位移校正的Imagika和Charnley Duo方法。使用了5种具有不同假体组件的定制可调节体模。结果:在模拟磨损的三个级别(0.2毫米,1.0毫米和1.5毫米)下,对所有5个体模进行20次测量,数字RSA检查的平均测量误差为0.010毫米(精度为0.42)。三种射线照相技术的相应误差值分别是:Charnley Duo为0.19(精度1.3),校正后的Imagika为0.13(精度1.3)和Imagika为1.021(精度2.99)。测量误差从普通RSA的0.011 mm降低到RSA数字测量的0.004。头部尺寸,相对于罩杯的佩戴方向或假体类型不影响测量误差。 Charnley Duo和Imagika校正后的结果相似,但后者在测量幻像之一时存在无法解释的系统误差。由于无法补偿头顶中心效应,Imagika的结果最差。氧化铝头很难用所有方法进行分析。解释:通过使用ISO标准评估准确性,可以预期RSA能够以大约0.4毫米的精度测量磨损,而与所研究的假体部件或磨损方向无关,而我们研究中基于标准射线照片的最佳技术是精确到约1.3毫米。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号