Biomechanical analysis of an interference screw and a novel twist lock screw design for bone graft fixation

机译:干涉螺钉的生物力学分析及骨移植固定的新型扭锁螺杆设计

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Abstract Background Malpositioning of an anterior cruciate ligament graft during reconstruction can occur during screw fixation. The purpose of this study is to compare the fixation biomechanics of a conventional interference screw with a novel Twist Lock Screw, a rectangular shaped locking screw that is designed to address limitations of graft positioning and tensioning. Methods Synthetic bone (10, 15, 20lb per cubic foot) were used simulating soft, moderate, and dense cancellous bone. Screw push-out and graft push-out tests were performed using conventional and twist lock screws. Maximum load and torque of insertion were measured. Findings Max load measured in screw push out with twist lock screw was 64%, 60%, 57% of that measured with conventional screw in soft, moderate and dense material, respectively. Twist lock max load was 78% and 82% of that with conventional screw in soft and moderate densities. In the highest bone density, max loads were comparable in the two systems. Torque of insertion with twist lock was significantly lower than with conventional interference screw. Interpretation Based on geometric consideration, the twist lock screw is expected to have 35% the holding power of a cylindrical screw. Yet, results indicate that holding power was greater than theoretical consideration, possibly due to lower friction and lower preloaded force. During graft push out in the densest material, comparable max loads were achieved with both systems, suggesting that fixation of higher density bone, which is observed in young athletes that require reconstruction, can be achieved with the twist lock screw. Highlights ? A conventional screw can cause host damage with interference graft fixation. ? Pre-tension of an anterior cruciate ligament graft is critical for fixation. ? Fixation with a novel screw design (Twist Lock) is compared to conventional screws. ? Twist lock screw required less torque than conventional screw for fixation. ? In higher bone densities, fixation with twist lock screw was comparable to that of conventional screw.
机译:在螺钉固定期间,在重建期间,在重建期间的前令韧带移植物的摘要背景可能发生。本研究的目的是将传统的干涉螺钉的固定生物力学与新型扭曲锁螺钉进行比较,矩形锁定螺钉,矩形锁定螺钉设计用于解决移植物定位和张紧的限制。方法使用合成骨(10,15,20脚)模拟柔软,中等和密集的松质骨。使用常规和扭转锁定螺钉进行螺旋推出和移植式推出测试。测量了最大载荷和插入的扭矩。调查结果在螺杆推出中测量的最大载荷分别用扭曲锁定螺杆测量为64%,60%,57%,其中常规螺杆分别以柔软的,中等和致密的材料为测量。扭锁最大载荷为78%和82%,其中常规螺杆柔软和中等密度。在最高的骨密度中,两个系统中最大载荷相当。用扭转锁插入的扭矩显着低于传统的干涉螺杆。基于几何考虑的解释,扭转锁定螺杆预计将具有35%的圆柱形螺钉的保持力。然而,结果表明,持有功率大于理论考虑,可能是由于摩擦较低和更低的预载力。在移植物中,在密度最密集材料中,两种系统实现了可比的最大载荷,表明在需要重建的年轻运动员中观察到更高密度骨骼的固定,可以通过扭转锁定螺钉来实现。强调 ?传统的螺钉可以引起干涉移植固定的宿主损坏。还前坩埚韧带移植物的预张力对于固定至关重要。还将用新型螺钉设计(扭锁)固定与传统螺钉相比。还扭转锁定螺杆需要比传统螺钉更少的扭矩进行固定。还在较高的骨密度中,用扭转锁螺杆固定与传统螺钉相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号