首页> 外文期刊>Catheterization and cardiovascular interventions: Official journal of the Society for Cardiac Angiography & Interventions >Pushing boundaries: Implantation of the 34 mm Medtronic CoreValve in patients with a large aortic annulus
【24h】

Pushing boundaries: Implantation of the 34 mm Medtronic CoreValve in patients with a large aortic annulus

机译:推行边界:植入大主动脉环患者34毫米Medtronic Corevalve

获取原文
获取原文并翻译 | 示例
       

摘要

The relationship between adherence to the recommended CoreValve sizing parameters and clinical outcomes is not well known for the recently released 34 mm valve, which is currently the largest available transcatheter valve. There is a presumed temporal reduction in paravalvular regurgitation in patients who receive an in-range valve, however, certain patients possess annular dimensions that are too large. We therefore describe two patients with annular dimensions larger than the manufacturer recommended range for the 34 mm CoreValve, who despite this underwent transcatheter aortic valve replacement with excellent clinical outcomes.
机译:在最近释放的34 mm阀门,遵守推荐的核心尺寸参数和临床结果之间的关系是目前最大的可用转阀阀门的关系。 在接受无线阀的患者中存在静脉瓣膜瓣膜瓣膜瓣膜的时间减少,然而,某些患者具有太大的环形尺寸。 因此,我们描述了两种环形尺寸的患者,这是制造商推荐的34 mm CoreValve的范围,尽管这种经过经过经过的经膜状管主动脉瓣膜,但具有优异的临床结果。