Monte Carlo dosimetric characterization of the Flexisource Co-60 high-dose-rate brachytherapy source using PENELOPE

机译:Monte Carlo Dosimetric使用Penelope的Flexisource Co-60高剂量速率近距离放射治疗源

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Co-60 sources have been commercialized as an alternative to Ir-192 sources for high dose-rate (HDR) brachytherapy. One of them is the Flexisource Co-60 HDR source manufactured by Elekta. The only available dosimetric characterization of this source is that of Vijande et al. [J Contemp Brachytherapy 2012; 4:34-44], whose results were not included in the AAPM/ESTRO consensus document. In that work, the dosimetric quantities were calculated as averages of the results obtained with the Geant4 and PENELOPE Monte Carlo (MC) codes, though for other sources, significant differences have been quoted between the values obtained with these two codes. The aim of this work is to perform the dosimetric characterization of the Flexisource Co-60 HDR source using PENELOPE.
机译:目的:已将CO-60来源商业化为IR-192的替代品,用于高剂量率(HDR)近距离放射治疗。 其中一个是Elekta制造的FlexiSource Co-60 HDR源。 这个来源的唯一可用的描述是Vijande等人。 [J Contemp Brachytherapy 2012; 4:34-44],其结果不包括在AAPM / ESTRO共识文件中。 在该工作中,计算剂量计量计算为使用GEANT4和Penelope Monte Carlo(MC)代码获得的结果的平均值,但对于其他来源,在用这两个代码获得的值之间引用了显着的差异。 这项工作的目的是使用Penelope执行Flexisource Co-60 HDR源的描述。