Evaluation of Efficacy of White Blood Cell Identification in Peripheral Blood by Automated Scanning of Stained Blood Smear Images with Variable Magnification

机译:通过可变倍率自动扫描通过自动扫描评价外周血白细胞鉴定的疗效

获取原文
获取原文并翻译 | 示例
           

摘要

Development of blood smear scanner analyzers is a new promising trend in designing devices for blood cell morphology examination. The efficacy of using the Vision Hema Ultimate scanner analyzer for differential white blood cell count and identification of nucleated blood cells was evaluated. The average time required to analyze a single cover-glass preparation was found to be (108 ± 17)". There was no statistically significant difference between the results of differential white blood cell count performed manually and using the Vision Hema Ultimate device (p >0.05); the results were virtually identical. The Spearman's correlation coefficient was highest for segmented neutrophils, lymphocytes and eosinophils (0.9638%, 0.9342%, and 0.9172%, respectively). For monocytes and basophils it was 0.9047% and 0.7613%, respectively, while its lowest values were observed for immature myeloid cells. The Vision Hema Ultimate blood smear scanner analyzer was shown to provide high accuracy and productivity of differential white blood cell count in peripheral blood smears.
机译:血液涂抹扫描仪分析仪的发展是一种新的血液细胞形态学考试设备的有前景趋势。评估使用视觉HEMA终极扫描仪分析仪进行差异白细胞计数的疗效和核肉细胞鉴定。发现分析单个覆盖玻璃制剂所需的平均时间是(108±17)“。手动进行的差异白细胞计数的结果与使用视觉HEMA极限装置(P>)没有统计学上显着差异(P> 0.05);结果几乎是相同的。分段中性粒细胞,淋巴细胞和嗜酸性粒细胞(0.9638%,0.9342%,分别为0.9172%,Spearman的相关系数最高。对于单核细胞和嗜碱性粒细胞,分别为0.9047%和0.7613%,对于未成熟的髓样细胞,观察到其最低值。视觉HEMA终极血液涂抹扫描仪分析仪在外周血涂片中提供差异白细胞计数的高精度和生产率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号