Validation of an automatic device for the self-measurement of blood pressure in sitting and supine positions according to the ANSI/AAMI/ISO81060-2:2013 guidelines: the Omron HEM-9700T

机译:根据ANSI / AAMI / ISO81060-2:2013指南,验证坐姿和仰卧位的自动测量血压的自动测量和仰卧位:OMRON HEM-9700T

获取原文
获取原文并翻译 | 示例
           

摘要

Aim We evaluated the performance of the Omron HEM-9700T, an automatic upper-arm-type device for the self-measurement of blood pressure (BP), in the sitting and supine positions according to the ANSI/AAMI/ISO81060-2:2013 guidelines. Participants and methods We screened 106 participants for the validation study in both positions. After excluding 21 participants, a total of 85 participants [38 (44.7%) men; 47 (55.3%) women] with mean +/- SD age of 54.5 +/- 12.2 years (range: 21-78 years) were included. Their arm circumference was 27.5 +/- 1.4cm (range: 17.9-40.8cm). The arm circumference distribution met the ANSI/AAMI/ISO81060-2:2013 requirement. The participants' BP measurements alternated between the use of a conventional mercury sphygmomanometer and the automatic device according to the ANSI/AAMI/ISO81060-2:2013 guidelines to measure their BP. Results The mean differences between reference BPs and HEM-9700T readings in the sitting and supine positions were -0.6 +/- 6.4/1.3 +/- 5.5 and -1.0 +/- 6.2/1.9 +/- 5.4mmHg for systolic BP/diastolic BP, respectively. Conclusion The Omron HEM-9700T fulfilled the validation criteria of the ANSI/AAMI/ISO81060-2:2013 guidelines for both the sitting and supine positions. (C) 2019 Wolters Kluwer Health, Inc. All rights reserved.
机译:目的我们评估了欧姆龙HEM-9700T的性能,自动上臂式装置,用于自动测量血压(BP),坐姿和仰卧位根据ANSI / AAMI / ISO8106060-2:2013指导方针。参与者和方法我们在两个职位上筛选了106名参与者的验证研究。排除21名参与者后,共有85名参与者[38(44.7%); 47(55.3%)妇女]平均+/- SD年龄为54.5 +/- 12.2岁(范围:21-78岁)。他们的臂周长为27.5 +/- 1.4cm(范围:17.9-40.8cm)。 ARM周长分布符合ANSI / AAMI / ISO81060-2:2013要求。参与者的BP测量在使用传统的汞血压计和根据ANSI / AAMI / ISO81060-2:2013指南的使用之间交替使用,以衡量其BP。结果坐姿和仰卧位中参考BPS和下摆9700T读数之间的平均差异为-0.6 +/- 6.4 / 1.3 +/- 5.5和-1.0 +/- 6.2 / 1.9 +/- 5.4mmHg,用于收缩性BP /舒张分别为BP。结论Omron HEM-9700T满足了ANSI / AAMI / ISO81060-2:2013年坐姿和仰卧位指南的验证标准。 (c)2019 Wolters Kluwer Health,Inc。保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号