The estimation scheme of uncertainty of determination of 1-hydroxypyrene (1-OHP) in urine was developed analysing the main stages of the analytical procedure: (1) preparation of 1-OHP standards, (2) creation of the calibration curve for the high performance liquid chromatography (HPLC) analysis method with the evaluation of recovery, (3) measuring procedure of aliquot of urine, (4) adjusting the pH of aliquot and hydrolysis with enzyme, (5) solid phase extraction, (6) concentration of the extract, (7) injection of the extract to chromatograph and analysing by the HPLC method, (8) calculation of 1-OHP mass from the calibration curve, (9) calculation of 1-OHP concentration in urine. The evaluation of the uncertainty is based on quantification of individual components. Combined uncertainty was calculated using the law of propagation of uncertainties according to the EURACHEM/CITAC guidelines. Level dependence of the uncertainty arises from the calibration curve. The limits of detection and quantification were found to be equal to 0.03 and 0.1 ng/mL, respectively. The calculated expanded level-dependent uncertainty covers 47-27-25% within the concentration range 0.03-0.1-0.4 ng/mL with the materials and equipment used. These parameters could easily be recalculated according to the proposed scheme if there are some changes in the analysis procedure.
机译:通过分析分析程序的主要阶段,制定了尿中1-羟基py(1-OHP)测定不确定度的估算方案:(1)制备1-OHP标准品;(2)建立高浓度标定曲线高效液相色谱(HPLC)分析方法,评估回收率;(3)测定尿液的等分步骤;(4)调节等分试样的pH值并用酶水解;(5)固相萃取;(6)提取物,(7)将提取物注入色谱仪并通过HPLC方法进行分析,(8)根据校准曲线计算1-OHP质量,(9)计算尿液中1-OHP浓度。不确定性的评估基于单个组件的量化。根据EURACHEM / CITAC准则,使用不确定性传播定律计算组合的不确定性。不确定性的水平依赖性源自校准曲线。检出限和定量限分别等于0.03和0.1 ng / mL。在0.03-0.1-0.4 ng / mL的浓度范围内,使用所使用的材料和设备,计算得出的扩展的取决于水平的不确定度涵盖47-27-25%。如果分析过程中有一些变化,可以很容易地根据所提出的方案重新计算这些参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号