Parallel placement of Excluder legs for treatment of type IIIb endoleaks caused by fabric tear after endovascular aneurysm repair

机译:用于治疗血管内动脉瘤修复后的织物撕裂型IIIB末端的护排腿的并行放置

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A total of 576 patients underwent endovascular aneurysm repair using main body devices for treatment of abdominal aortic aneurysms or iliac artery aneurysms. During follow-up, type IIIb endoleaks caused by fabric tear occurred in six patients (1.0% [6/576]). The device used was Zenith (Cook Medical, Bloomington, Ind) in five cases and Talent (Medtronic, Santa Rosa, Calif) in one case. All endoleaks were close to the flow divider of the main body devices. The distance between the lower renal artery and the top end of the contralateral leg was 53 +/- 14 mm. Bell-bottom-shaped Excluder (W. L. Gore & Associates, Flagstaff, Ariz) legs were placed parallel from the top of the main body device through both legs to treat these endoleaks. In two patients, coil embolization was required to treat gutter endoleaks. Postoperative computed tomography showed the obliteration of type IIIb endoleaks in all patients. Our technique may be an acceptable method for treatment of type IIIb endoleaks, especially when they occur near the flow divider.
机译:共有576例患者使用用于治疗腹主动脉瘤或髂动脉瘤的主体装置进行血管内动脉瘤修复。在随访期间,六名患者发生织物撕裂型IIIB末端(1.0%[6/576])。使用的设备是Zenith(Cook Medical,Bloomington,Ind)在五种情况下(Medtronic,Santa Rosa,Calif)在一个案例中。所有腹部均靠近主体装置的流量分压器。下肾动脉和对侧腿的顶端之间的距离为53 +/- 14毫米。贝尔底部形状的排斥(W.L. Gore&Associates,Flagstaff,Ariz)腿通过双腿从主体装置的顶部平行地放置,以处理这些螺旋凸。在两名患者中,需要卷曲栓塞来治疗沟槽螺旋疱疹。术后计算断层扫描显示所有患者IIIB型胚胎的爆炸。我们的技术可以是用于治疗IIIB型腹腔的可接受的方法,尤其是当它们发生在流动分隔附近时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号