Hi'/> Column-to-column packing variation of disposable pre-packed columns for protein chromatography
首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >Column-to-column packing variation of disposable pre-packed columns for protein chromatography
【24h】

Column-to-column packing variation of disposable pre-packed columns for protein chromatography

机译:用于蛋白质色谱的一次性预填充柱的柱到柱填充变化

获取原文
获取原文并翻译 | 示例
           

摘要

Highlights?The column-to-column variation mainly affects column efficiency, not the retention.?The column efficiency varies between 10 and 15% for equally packed columns.?The column-to-column variation is not explained by the packing density.?The column-to-column variation is explained by heterogeneity in packing structure.?Measurement precision of the workstation was included in the evaluation.AbstractIn the biopharmaceutical industry, pre-packed columns are the standard for process development, but they must be qualified before use in experimental studies to confirm the required performance of the packed bed. Column qualification is commonly done by pulse response experiments and depends highly on the experimental testing conditions. Additionally, the peak analysis method, the variation in the
机译:<![cdata [ 突出显示 列到列的变化主要影响列效率,而不是保留。 列效率在10到15%之间变化为10%和15%同样打包的列。 填充密度不解释列到列变型。 标签> 列到列变量是通过包装结构中的异质性解释。 计算工作站的测量精度包含在评估中。 < / ce:abstract-sec> 抽象 在生物制药行业中,预包列列是过程开发的标准,但他们之前必须有资格用于实验研究以确认填充床的所需性能。柱鉴定通常通过脉冲响应实验完成,并在高度上依赖于实验测试条件。另外,峰值分析方法,变化