Binding constants determination of cyclodextrin inclusion complexes by affinity capillary electrophoresis. How to overcome the limitations induced by the UV-detector?

机译:结合常数通过亲和毛细管电泳测定环糊精包合物。 如何克服UV检测器引起的局限性?

获取原文
获取原文并翻译 | 示例
           

摘要

In most cases, determination of binding constant for analyte-cyclodextrin complexes in capillary electrophoresis is investigated by affinity capillary electrophoresis using a UV detector (ACE-UV). The limitations induced by the UV-detector include : (i) the difficulty of dealing with poor chromogenic analytes and more generally with any analyte presenting strong affinity towards the cyclodextrin (CD), i.e. for which the prerequisite to work with analyte concentration much smaller than those of the CD is difficult to fulfill (ii) the impossibility of studying non-chromogenic analyte. In this paper, two simple methodologies were developed to overcome these limitations. Regarding the analytes which present poor UV-absorbance and/or very high CD-affinity, a methodology using an algorithmic data treatment and taking into account the real analyte concentration in the capillary at the determined migration times allows to correctly estimate the binding constants, even if the experimental prerequisite ([analyte][CD]) is not complied. Moreover, it is proved that classical linearization treatment by picking the migration time of the infinite diluted analytes (at the start of the peak) also provide satisfactory results. Regarding UV-transparent analyte, a competitive methodology combined with algorithmic data treatment allows the determination of their affinity towards cyclodextrins. Last, the applicability of the described competitive method is extended to the study of interaction between two neutral partners, which is another well-known limitation of ACE. (C) 2020 Published by Elsevier B.V.
机译:在大多数情况下,通过使用UV检测器(ACE-UV),通过亲和毛细管电泳研究毛细管电泳中分析物 - 环糊精复合物的结合常数。紫外检测器诱导的局限性包括:(i)难以处理差的发色分析物,更通常用朝向环糊精(CD)具有强亲和力的任何分析物,即,使用远小于的分析物浓度的先决条件CD的那些难以实现(ii)研究非显色分析物的不可能性。在本文中,开发了两种简单的方法来克服这些限制。关于紫外光吸收和/或非常高的CD-亲和力的分析物,使用算法数据处理的方法,并考虑确定的迁移时间在毛细管中的真实分析物浓度允许正确估计结合常数,甚至如果实验前提条件([分析物] [CD])不遵守。此外,证明了通过挑选无限稀释的分析物的迁移时间(在峰的开始时)的典型线性化处理也提供了令人满意的结果。关于UV透明分析物,与算法数据处理相结合的竞争方法允许测定它们对环糊精的亲和力。最后,所描述的竞争方法的适用性延伸到两个中立伙伴之间的相互作用的研究,这是ACE的另一个知名限制。 (c)2020由elsevier b.v发布。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号