Near-infrared fluorescence laparoscopy of the ureter with three preclinical dyes in a pig model

机译:猪水模型中具有三种临床前染料的近红外荧光腹腔镜检查

获取原文
获取原文并翻译 | 示例

摘要

BackgroundUreteric injury is reported to occur in 1-7.6% of colorectal surgeries. To reduce the incidence of ureteral injury, it is essential to identify the ureters. The use of near-infrared fluorescence (NIRF) imaging with intravenously administered dyes might be of added value for ureteral visualization during laparoscopy. The aim of this study is to assess the performance of three preclinical dyes; IRDye((R)) 800BK, IRDye((R)) 800NOS and IRDye((R)) 800CW, for near-infrared fluorescence laparoscopy of the ureter in pigs.MethodsIn three female Dutch landrace pigs, the new dyes were evaluated. In each pig, 1 dye was tested using a 6-mg intravenous dose in a concentration of 1mg/ml. Imaging was performed in fluorescence mode and white light mode with a laparoscopic imaging system. In order to further evaluate the dyes, an ex vivo imaging experiment was performed, in which 8 decreasing concentrations per dye, diluted in PBS, were evaluated in a transparent test tube with NIRF mode at a distance of 1, 5 and 10cm from the laparoscope.ResultsAll three dyes were effective in allowing the identification of the ureter with NIRF imaging. The ureter became fluorescent after 35, 45 and 10min, respectively, for IRDye((R)) 800BK, IRDye((R)) 800NOS and IRDye((R)) 800CW with a maximum target-to-background ratio (TBR) of 2.14, 0.66 and 1.44, respectively. In the ex vivo imaging experiment, all three dyes produced a strong fluorescence signal at all concentrations and all distances evaluated.ConclusionsIntravenous administration of the preclinical dyes IRDye((R)) 800CW, IRDye((R)) 800 BK and IRDye((R)) 800NOS facilitated successful identification of the anatomical course of the ureter in living pig models. The highest measured TBR occurred with the use of IRDye((R)) 800BK. Ex vivo, a correlation was observed between the fluorescence intensities of the signal with the concentration of the dye and with the distance to the object.
机译:据报道,背景伤害以1-7.6%的结肠直肠手术。为了减少输尿管损伤的发生率,必须识别输尿管。使用近红外荧光(NIRF)成像与静脉内给药染料的成像可能是腹腔镜检查期间输尿管可视化的附加值。本研究的目的是评估三种临床前染料的性能; IRDYE((r))800bk,irdye((r))800nos和irdye((r))800cw,用于猪疹的近红外荧光腹腔镜检查。方法三个雌性荷兰地区猪,评估了新的染料。在每只猪中,使用6mg静脉注射剂量的浓度为1mg / ml来测试1染料。用腹腔镜成像系统以荧光模式和白光模式进行成像。为了进一步评价染料,进行了前体内成像实验,其中在从腹腔镜的距离为1,5和10cm的透明试管中评估在PBS中稀释的每染料的80染料的减少。 .Resultsall三种染料在允许用NIRF成像鉴定输尿管。在35,45和10min的情况下,对于IRDYE((r))800bk,irdye((r))800nos和irdye((r))800cw,具有最大目标 - 背景比(TBR),输尿管率分别变为荧光。 2.14,0.66和1.44分别。在前体内成像实验中,所有三种染料在所有浓度下产生强大的荧光信号,并达到所有距离。结合临床前染料IRDYE((R))800CW,IRDYE((R))800 BK和IRDYE((r ))800NOS促进了生活猪模型中输尿管的解剖程序的成功鉴定。使用IRDYE((R))800BK的使用最高测量的TBR。例如,在信号的荧光强度与染料浓度和与物体的距离之间观察到相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号