摘要

In the case of clinical ventilation, sensors are needed that analyze the carbon dioxide concentration during the exhalation cycle. For devices capable to individually control and adjust parameters for each patient, the analysis of the breathing conditions including carbon dioxide concentration is of critical importance. We present a miniaturized system that is optimized for a fast passive gas exchange with low dead volumes in the side stream. The presented system which is also suitable for neonates shows high dynamics and no influence on the breathing behavior due to small systems sizes. Cost reduction is achieved through a fully passive gas flow that does not require active elements such as valves and pumps. The cuvette has been designed for optimized optical and fluidic properties for surveillance in a wide range of breathing conditions. (C) 2018 Elsevier B.V. All rights reserved.
机译:在临床通气的情况下,需要传感器,以在呼出循环期间分析二氧化碳浓度。 对于能够单独控制和调节每位患者的参数的装置,对包括二氧化碳浓度的呼吸条件的分析具有至关重要的重要性。 我们提出了一种小型化系统,该系统针对侧坯料中的低死体积进行了优化的优化。 由于小系统尺寸,所呈现的系统也适用于新生儿的系统,显示出高动态,对呼吸行为没有影响。 通过完全被动气体流动实现成本降低,这些气流不需要诸如阀门和泵等活动元件。 比色皿专为优化的光学和流体性能而设计,用于在各种呼吸条件下进行监视。 (c)2018年elestvier b.v.保留所有权利。