Technical Note: Quantitative accuracy evaluation for spectral images from a detector‐based spectral CT CT scanner using an iodine phantom

机译:技术说明:使用碘幻像的基于探测器的光谱CT CT CT扫描仪的光谱图像定量准确度评价

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose The purpose of this study was to evaluate the quantitative accuracy of spectral images from a detector‐based spectral CT scanner using a phantom with iodine‐loaded inserts. Methods A 40‐cm long‐body phantom with seven iodine inserts (2–20 mg/ml of iodine) was used in the study. The inserts could be placed at 5.5 or 10.5 cm from the phantom axis. The phantom was scanned five times for each insert configuration using 120 kV p tube voltage. A set of iodine, virtual noncontrast, effective atomic number, and virtual monoenergetic spectral CT images were generated and measurements were made for all the iodine rods. Measured values were compared with reference values calculated from the chemical composition information provided by the phantom manufacturer. Radiation dose from the spectral CT was compared to a conventional CT using a CTDI (32 cm) phantom. Results Good agreement between measurements and reference values was achieved for all types of spectral images. The differences ranged from ?0.46 to 0.1 mg/ml for iodine concentration, ?9.95 to 6.41 HU for virtual noncontrast images, 0.12 to 0.35 for effective Z images, and ?17.7 to 55.7 HU for virtual monoenergetic images. For a similar CTDI vol, image noise from the conventional CT was 10% lower than the spectral CT . Conclusions The detector‐based spectral CT can achieve accurate spectral measurements on iodine concentration, virtual non‐contrast images, effective atomic numbers, and virtual monoenergetic images.
机译:目的本研究的目的是使用幻象与碘加载的刀片,以评估从基于检测器光谱CT扫描仪分光图像的定量精度。方法将40厘米长体幻七个碘插入物(2-20毫克/毫升的碘)中的溶液在研究中使用。插入件可以放置在距幻象轴5.5或10.5公分。所述体模进行5次扫描,使用120千伏p管电压的每个的插入配置。产生一组碘,虚拟非增强,有效原子序数,和虚拟单能的光谱CT图像并测量了对所有的碘杆制成。测量值与从由虚线制造商提供的化学成分的信息计算的参考值进行比较。从光谱CT辐射剂量与使用CTDI(32厘米)的虚线的常规CT。测量结果和参考值之间良好的一致性是为所有类型的光谱图像的实现。的差异的范围为?0.46至0.1mg / ml的碘浓度,?9.95到6.41 HU为虚拟非增强图像,0.12〜0.35为有效Ž图像,以及用于虚拟单能的图像?17.7至55.7 HU。对于类似CTDI体积,从常规的CT图像噪声为10%,比频谱CT降低。结论基于检测器光谱CT可以实现上碘浓度,虚拟非对比图像,有效原子序数,和虚拟单能准确的图像的光谱测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号