Frame average optimization of cine-mode EPID images used for routine clinical in vivo patient dose verification of VMAT deliveries

机译:用于常规临床的Cine-Mode EPID图像的帧平均优化在VMAT交付的体内患者剂量验证中使用

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: The in vivo 3D dose delivered to a patient during volumetric modulated arc therapy (VMAT) delivery can be calculated using electronic portal imaging device (EPID) images. These images must be acquired in cine-mode (i.e., "movie" mode) in order to capture the time-dependent delivery information. The angle subtended by each cine-mode EPID image during an arc can be changed via the frame averaging number selected within the image acquisition software. A large frame average number will decrease the EPID's angular resolution and will result in a decrease in the accuracy of the dose information contained within each image. Alternatively, less EPID images acquired per delivery will decrease the overall 3D patient dose calculation time, which is appealing for large-scale clinical implementation. Therefore, the purpose of this study was to determine the optimal frame average value per EPID image, defined as the highest frame averaging that can be used without an appreciable loss in 3D dose reconstruction accuracy for VMAT treatments.
机译:目的:可以使用电子门户成像装置(EPID)图像计算在体积调制的电弧疗法(VMAT)输送期间递送到患者的体内3D剂量。必须以Cine-Mode(即,“电影”模式)以Cine-Mode(即“)获取这些图像,以捕获时间相关的传送信息。可以通过图像获取软件中选择的帧平均数来改变由弧期间的每个Cine模式ePID图像下调的角度。大帧平均数将降低EPID的角度分辨率,并且将导致每个图像内包含的剂量信息的准确性降低。或者,每递送获取的较少的EPID图像将降低整个3D患者剂量计算时间,这是对大规模临床实现的吸引力。因此,本研究的目的是确定每个EPID图像的最佳帧平均值,定义为可以使用的最高帧平均,而没有用于VMAT处理的3D剂量重建精度的明显损失。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号