首页> 外文期刊>European journal of cardio-thoracic surgery: Official journal of the European Association for Cardio-thoracic Surgery >Stent grafting simulation using a three-dimensional printed model for extensive aortic arch repair combined with coarctation
【24h】

Stent grafting simulation using a three-dimensional printed model for extensive aortic arch repair combined with coarctation

机译:使用三维印刷模型的支架仿真模拟,用于广泛的主动脉弓修复结合凸起

获取原文
获取原文并翻译 | 示例

摘要

Stent grafting for complex aortic anatomy remains a challenge. In particular, stent graft collapse (i.e. infolding) is possible when an excessive oversized device is needed. We describe a case of preoperative stent grafting simulation using a three-dimensional (3D) printed model for extensive aortic arch repair in a 69-year-old woman with multiple aneurysms combined with coarctation. The patient was scheduled to undergo staged hybrid repair. A stent graft larger than 28mm in diameter was needed to deploy into a coarctation of 15mm in diameter during the 2nd stage of the operation. Preoperative, experimental stent grafting using a 3D printed model indicated that infolding would likely not occur. Therefore, we proceeded with surgery, which was successful. This technology could be a useful application for planning complicated stent grafting.
机译:复杂主动脉解剖学支架嫁接仍然是一项挑战。 特别是,当需要过大的过大的装置时,可以使用支架移植物折叠(即infOSDING)。 我们描述了一种使用三维(3D)印刷模型的术前支架仿真的案例,该模型在一名69岁的女性中,具有多个动脉瘤联合Cyctication的一个69岁的女性。 患者计划进行分阶段的混合修复。 需要大于28mm的支架移植物,以在操作的第二阶段展开直径15mm的凸起。 使用3D打印模型的术前,实验支架接枝表明,可能不会发生令人反感。 因此,我们进行手术,这是成功的。 这项技术可能是规划复杂支架嫁接的有用应用程序。