High-performance liquid chromatography coupled to direct analysis in real time mass spectrometry: Investigations on gradient elution and influence of complex matrices on signal intensities

机译:高效液相色谱结合实时质谱法中的直接分析:梯度洗脱以及复杂基质对信号强度的影响的研究

获取原文
获取原文并翻译 | 示例
           

摘要

Direct analysis in real time (DART) time-of-flight mass spectrometry (TOF-MS) has been tested for its suitability as a detector for gradient elution HPLC. Thereby a strong dependency of signal intensity on the amount of organic solvent present in the eluent could be observed. Adding a make-up liquid (iso-propanol) post-column to the HPLC effluent greatly enhanced detection limits for early eluting compounds. Limits of detection achieved employing this approach were in the range of 7-27μgL~(-1) for the parabene test mixture and 15-87μgL~(-1) for the pharmaceuticals. In further investigations DART ionization was compared to several other widely used atmospheric pressure ionization methods with respect to signal suppression phenomena occurring in when samples with problematic matrices are analyzed. For this purpose extracts from environmental and waste water samples were selected as model matrices which were subsequently spiked with a set of six substances commonly present in personal care products as well as six pharmaceuticals at concentration levels between 100μgL~(-1) and 500μgL~(-1) corresponding to 100ngL~(-1) and 500ngL~(-1) respectively in the original sample. With ionization suppression of less than 11% for most analytes investigated, DART ionization showed similar to even somewhat superior behavior compared to atmospheric pressure chemical ionization (APCI) and atmospheric pressure photo ionization (APPI) for the Danube river water extract; for the more challenging matrix of the sewage plant effluent extract DART provided better results with ion suppression being less than 11% for 9 out of 12 analytes while values for APCI were lying between 20% and >90%. Electrospray ionization (ESI) was much more affected by suppression effects than DART with values between 26% and 80% for Danube river water; in combination with the sewage plant effluent matrix suppression >50% was observed for all analytes.
机译:实时直接分析(DART)飞行时间质谱(TOF-MS)已作为梯度洗脱HPLC检测器进行了测试。因此,可以观察到信号强度对洗脱液中有机溶剂含量的强烈依赖性。色谱柱后添加补充液(异丙醇)到HPLC流出物中,大大提高了早期洗脱化合物的检测限。采用这种方法所达到的检出限在抛物线测试混合物中为7-27μgL〜(-1),在药物中为15-87μgL〜(-1)。在进一步的研究中,将DART电离与其他几种广泛使用的大气压电离方法进行了比较,以解决在分析有问题的基质样品时发生的信号抑制现象。为此,从环境和废水样品中选择提取物作为模型基质,然后将其掺入一组个人护理产品中常见的六种物质以及六种药物,浓度范围为100μgL〜(-1)至500μgL〜( -1)分别对应于原始样本中的100ngL〜(-1)和500ngL〜(-1)。大多数被分析物的电离抑制低于11%,与多瑙河水提取物的大气压化学电离(APCI)和大气压光电离(APPI)相比,DART电离表现出甚至更高的性能。对于更具挑战性的污水处理厂废水萃取物基质,DART提供了更好的结果,其中12种分析物中的9种离子抑制低于11%,而APCI值在20%至> 90%之间。与DART相比,电喷雾电离(ESI)受抑制作用的影响要大得多,多瑙河水的浓度在26%至80%之间。结合污水处理厂废水基质,所有分析物的抑制均> 50%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号