首页> 外文期刊>Journal of chromatography, A: Including electrophoresis and other separation methods >The design and use of a simple System Suitability Test Mix for generic reverse phase high performance liquid chromatography-mass spectrometry systems and the implications for automated system monitoring using global software tracking
【24h】

The design and use of a simple System Suitability Test Mix for generic reverse phase high performance liquid chromatography-mass spectrometry systems and the implications for automated system monitoring using global software tracking

机译:用于通用反相高效液相色谱-质谱系统的简单系统适用性测试混合物的设计和使用,以及使用全局软件跟踪进行自动化系统监控的意义

获取原文
获取原文并翻译 | 示例
           

摘要

The development of a seven-component test mixture designed for use with a generic gradient and a reversed-phase high performance liquid chromatography-mass spectrometry (RP-HPLC-MS) system is discussed. Unlike many test mixtures formulated in order to characterise column quality at neutral pH, the test mixture reported here was designed to permit an overall suitability assessment of the whole liquid chromatography-mass spectrometry (LCMS) system. The mixture is designed to test the chromatographic performance of the column as well as certain aspects of the performance of the individual instrumental components of the system. The System Suitability Test Mix can be used for low and high pH generic reverse phase LCMS analysis. Four phthalates are used: diethyl phthalate (DEP), diamyl phthalate (DAP), di-n-hexyl phthalate (DHP) and dioctyl phthalate (DOP). Three other probes are employed: 8-bromoguanosine (8-BG), amitryptyline (Ami), and 4-chlorocinnamic acid (4-CCA). We show that analysis of this test mixture can alert the user when any part of the system (instrument or column) contributes to loss of overall performance and may require remedial action and demonstrate that it can provide information that enables us to document data quality control.
机译:讨论了设计用于通用梯度和反相高效液相色谱-质谱(RP-HPLC-MS)系统的七组分测试混合物的开发。与为了在中性pH下表征色谱柱质量而配制的许多测试混合物不同,此处报告的测试混合物旨在对整个液相色谱-质谱(LCMS)系统进行总体适用性评估。该混合物旨在测试色谱柱的色谱性能以及系统中各个仪器组件的某些性能方面。系统适用性测试混合物可用于低和高pH通用反相LCMS分析。使用了四种邻苯二甲酸酯:邻苯二甲酸二乙酯(DEP),邻苯二甲酸二戊酯(DAP),邻苯二甲酸二正己酯(DHP)和邻苯二甲酸二辛酯(DOP)。使用了其他三种探针:8-溴鸟苷(8-BG),阿米替林(Ami)和4-氯肉桂酸(4-CCA)。我们表明,当系统的任何部分(仪器或色谱柱)造成整体性能下降并且可能需要采取补救措施时,对此测试混合物的分析都可以提醒用户,并证明它可以提供使我们能够记录数据质量控制的信息。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号