123I-MIBG Cardiac Imaging: Acquisition Protocols and Correction Methods for Quantitative Evaluation

机译:123I-MIBG心脏成像:定量评估的采集方案和校正方法

获取原文
获取原文并翻译 | 示例
           

摘要

Cardiac sympathetic imaging with 123I-MIBG offers unique insights into heart failure and neurodegenerative disease. Because of the diffuse nature of impairment, quantitative evaluation is indispensable, and the heart-to-mediastinum ratio is widely accepted as a valuable index. However, estimation of this index is highly dependent on acquisition protocol, especially collimator choice. A low-energy high-resolution collimator having thin septa is often used for 123I-MIBG imaging, and high-energy photons emitted from 123I readily penetrate the thin septa. The heart-to-mediastinum ratios are affected by accumulation in the lung and liver and are underestimated to various degrees. Medium-energy collimators having thick septa are less susceptible to septal penetration and are expected to offer a better indicator of cardiac sympathetic function. This review will provide an overview of the physics of 123I imaging and discuss acquisition protocols and correction methods for quantitative evaluation of cardiac 123I-MIBG imaging with special reference to the effect of septal penetration.
机译:123I-MIBG进行的心脏交感成像可对心力衰竭和神经退行性疾病提供独特见解。由于损伤的分散性,定量评估是必不可少的,并且心脏纵隔比率被广泛接受为有价值的指标。但是,该指数的估计高度依赖于采集协议,尤其是准直仪的选择。具有薄间隔的低能量高分辨率准直仪通常用于123I-MIBG成像,并且从123I发射的高能量光子容易穿透薄间隔。心脏与纵隔的比率受肺和肝中积累的影响,并在不同程度上被低估。隔垫较厚的中能准直器不易穿透隔垫,有望提供更好的心脏交感功能指标。这篇综述将概述123I成像的物理原理,并讨论对心脏123I-MIBG成像进行定量评估的采集方案和校正方法,并特别参考间隔穿透的效果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号